Drugs & medicalClass II

Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 64 System Number: 882471 Continuous CT (CCT) is a CT or…

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or…

Read the official notice →
Date
2021-12-01
Country
United States
Brand
Philips
Product
Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 64 System Number: 882471 Continuous CT (CCT) is a CT or…
Classification
Class II
Units
US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips recalls

  1. 2022-01-19
  2. US
    Drugs & medicalClass II

    Philips North America Llc recalls Wireless Footswtich used with the following Philips Allura Xper and Azurion Interventional Fluoroscopic X-ray System . Model Number:…

    Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Canada, Australia,…Official notice →
  3. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2…

    (1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide distribution in the state of WA and the…Official notice →
  4. 2021-12-29
  5. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular and neurovascular imaging…

    Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide distribution in the states of DC, GA,…Official notice →
  6. 2021-12-15
  7. US
    Drugs & medicalClass II

    Philips North America Llc recalls BV Endura with Software Release 2.3- A mobile, diagnostic X-ray image acquisition and viewing system. Model Number: 718075

    Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Afghanistan,…Official notice →
  8. 2021-11-17
  9. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078,…

    When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes…

    FDA · DevicePhilips North America LlcDomestic distribution nationwide. International distribution worldwide.Official notice →
  10. 2021-10-20
  11. US
    Drugs & medicalClass II

    Philips North America LLC recalls Philips, HeartStart HS1 OnSite Defibrillator, Model #'s: M5066A and M5067A

    Customers were not notified of previous recalls associated with various defibrillator models.

    FDA · DevicePhilips North America LLCU.S.: CA, CT, FL, GA, IN, MI, MO, NJ, NY, OH, TX, and WA. International…Official notice →
All 100 Philips recalls →

Other drugs & medical recalls in the United States

  1. 2021-12-01
  2. US
    Drugs & medicalClass II

    Medtronic Vascular, Inc. recalls Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE

    During stent graft deployment, the spindle may detach from the distal end of the spindle hypotube

    FDA · DeviceMedtronic Vascular, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of CA, FL,…Official notice →
  3. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls SOMATOM Definition Edge with software syngo.CT VB20 Model #10590000

    software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following…

    The processing of collected episode data may fail due to a software error.

    FDA · DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Worldwide distribution - US Nationwide. There was military/government…Official notice →
  5. US
    Drugs & medicalClass II

    Instrumentation Laboratory recalls HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

    The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or…

    FDA · DeviceInstrumentation LaboratoryWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  6. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls The da Vinci Xi and X surgical systems (IS4000 and IS4200) each consist of three main components: the Surgeon Console (SC), the Vision…

    Incorrect expiration date on its package labeling (shipper box and inner carton label).

    FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, CA,…Official notice →
  7. US
    Drugs & medicalClass II

    The Binding Site Group, Ltd. recalls Freelite Human Lambda Free Kit (for use on the SPAPLUS) REF LK018.S Freelite Human Lambda Free Kit (for use on the SPAPLUS) (x10)…

    Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6…

    FDA · DeviceThe Binding Site Group, Ltd.U.S.: CA, CO, FL, IL, IN, KY, LA, MI, NY, OK, SD, TX and VA. O.U.S.: CanadaOfficial notice →
See all drugs & medical recalls →