Olympus Corporation of the Americas recalls Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with…
Read the official notice →- Date
- 2022-06-22
- Country
- United States
- Brand
- Olympus Corporation of the Americas
- Product
- Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R
- Classification
- Class II
- Units
- CA FL MA MD MI NJ OH PR WI
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Olympus Corporation of the Americas recalls
- 2022-07-06
US
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model #…
Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.
FDA · DeviceOlympus Corporation of the AmericasNationwide distribution in the US. Foreign distribution worldwide.Official notice →- 2022-06-29
US
Olympus Corporation of the Americas recalls EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus,…
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of AL, FL, NC, NY.Official notice →US
Olympus Corporation of the Americas recalls Colonovideoscope Model CF-H180AL
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out…
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of KY.Official notice →- 2022-06-01
US
Olympus Corporation of the Americas recalls Olympus DISPOSABLE GRASPING FORCEPS, FG-51D
Forceps do not comply with Olympus standards for the amount of force required to open and close the forceps
- 2022-05-25
US
Olympus Corporation of the Americas recalls Olympus Uretero-reno videoscope, Model No. URF-V2
The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the state of AZ.Official notice →- 2022-05-18
US
Olympus Corporation of the Americas recalls BF-3C160: EVIS EXERA Bronchovideoscope
Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using…
Other drugs & medical recalls in the United States
- 2022-06-22
US
Volcano Corp recalls Philips OmniWire Pressure guide wire REF 89185J PN 300000252891
Due to a potential failed sterilization process.
US
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Crome Models loaded with CareLink SmartSync Device Manager application software D00U005: a) Crome VR:…
Telemetry error that may occur with Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
FDA · DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Distributed nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL GA, HI, IA,…Official notice →US
GE Healthcare, LLC recalls CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part…
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.
FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide. There was also military/government…Official notice →US
Micro Therapeutics, Inc. recalls AXIUM DETACHABLE COIL SYSTEM REF QC-4-12-HELIX; Axium Detachable Coil System REF QC-10-30-3D;
Due to incorrect size and configuration labeling of the detachable coil system.
FDA · DeviceMicro Therapeutics, Inc.International distribution in the countries of China and Republic of Korea.Official notice →US
Illumina, Inc. recalls illumina Model NextSeq 550 Dx REF 20005715
cybersecurity vulnerability
FDA · DeviceIllumina, Inc.Worldwide distribution - U.S. Nationwide distribution including in the states…Official notice →US
Tandem Diabetes Care Inc recalls t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology
Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will…