Drugs & medicalClass II

Olympus Corporation of the Americas recalls The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC…

The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body…

Read the official notice →
Date
2023-08-30
Country
United States
Brand
Olympus Corporation of the Americas
Product
The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC GASTROVIDEOSCOPE, ULTRASONIC…
Classification
Class II
Units
US Nationwide distribution.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Olympus Corporation of the Americas recalls

  1. 2023-09-06
  2. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Olympus Bronchovideoscope, Models BF-3C40 & BF-N20.

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Visera Hysterovideoscope Olympus HYF Type V

    IFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL,…Official notice →
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V

    Recent reports of patient infection.

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  5. 2023-08-23
  6. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls HF Resection Electrodes - HF resection electrodes are part of a resectoscope system for endoscopic diagnosis and treatment…

    Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls UroPass Ureteral Access Sheaths, 5 pieces/box

    Dilator tips may break in the package and in patients during surgical procedures.

    FDA · DeviceOlympus Corporation of the AmericasWorldwide distribution - United States Nationwide and the countries of South…Official notice →
  8. 2023-08-16
  9. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display…

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V…

    FDA · DeviceOlympus Corporation of the AmericasNationwideOfficial notice →
All 114 Olympus Corporation of the Americas recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-30
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the…

    There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA,…Official notice →
  3. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc.-Littleton recalls The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric…

    Ground cable installed incorrectly.

    FDA · DeviceMedtronic Navigation, Inc.-LittletonWorldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL,…Official notice →
  4. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…

    As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…

    FDA · DeviceRandox Laboratories Ltd.CA, FL, IL, ME, MI, NC, OH, PA,NJ,Official notice →
  5. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…

    (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…

    FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →
  6. US
    Drugs & medicalClass II

    Invacare Corporation recalls DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE…

    Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user

    FDA · DeviceInvacare CorporationUS Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail

    A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE…

    FDA · DeviceSmith & Nephew, Inc.US, AE, CA, DE, ES, GB, MX, PA, PLOfficial notice →