Drugs & medicalClass II

Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…

As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…

Read the official notice →
Date
2023-08-30
Country
United States
Brand
Randox Laboratories
Product
Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4,…
Classification
Class II
Units
CA, FL, IL, ME, MI, NC, OH, PA,NJ,
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Randox Laboratories recalls

  1. 2023-12-13
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.

    Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of AR, GA, NC, TN, & PR.Official notice →
  3. 2023-11-08
  4. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000

    RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution including Puerto RicoOfficial notice →
  5. 2023-09-27
  6. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215

    there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet…

    FDA · DeviceRandox Laboratories Ltd.US States: GA, OH, OR, PA, TX, VAOfficial notice →
  7. 2023-08-30
  8. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…

    (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…

    FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →
  9. 2023-08-16
  10. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)

    Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…Official notice →
  11. 2023-07-05
  12. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…

    Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…

    FDA · DeviceRandox Laboratories Ltd.CA, IL, VA, WA, WI including PROfficial notice →
All 27 Randox Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-30
  2. US
    Drugs & medicalClass II

    Medicrea International recalls IMPIX S, REF numbers: a) B15328071S, b) B15328073S, c) B15328074S, d) B15334072S, e) B15334073S, f) B15334074S; intervertebral fusion…

    There is a potential for a product packaging non-conformity issue which presents as a pinhole in either the inner or outer pouch.

    FDA · DeviceMedicrea InternationalUS nationwideOfficial notice →
  3. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the…

    There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA,…Official notice →
  4. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc.-Littleton recalls The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric…

    Ground cable installed incorrectly.

    FDA · DeviceMedtronic Navigation, Inc.-LittletonWorldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL,…Official notice →
  5. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC…

    The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…

    (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…

    FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →
  7. US
    Drugs & medicalClass II

    Invacare Corporation recalls DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE…

    Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user

    FDA · DeviceInvacare CorporationUS Nationwide distribution.Official notice →
See all drugs & medical recalls →