Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…
Read the official notice →- Date
- 2023-08-16
- Country
- United States
- Brand
- Randox Laboratories
- Product
- Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
- Classification
- Class II
- Units
- US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Randox Laboratories recalls
- 2023-09-27
US
Randox Laboratories Ltd. recalls Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215
there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet…
- 2023-08-30
US
Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…
As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…
US
Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…
FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →- 2023-07-05
US
Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…
- 2023-04-05
US
Randox Laboratories recalls CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826
The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
- 2023-01-11
US
Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical chemistry and…
Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.Official notice →
Other drugs & medical recalls in the United States
- 2023-08-16
US
Medtronic Neuromodulation recalls Medtronic InterStim Ground Pad REF 041826
Lot of InterStim Self-Adhesive Ground Pad is being recalled due to the product being labeled with the incorrect expiration date.
FDA · DeviceMedtronic NeuromodulationInternational distribution to the country of Iran.Official notice →US
Scholly Fiberoptic Gmbh recalls UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into…
There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this…
US
Medtronic Sofamor Danek USA, Inc recalls CD HORIZON Spinal System Solera Awl Tip Tap, REF NAV2006K
Incorrect Laser Marking - Solera Awl Tip Tap
FDA · DeviceMedtronic Sofamor Danek USA, IncUS Nationwide distribution in the states of FL, CO, IN, SC, TX, UT.Official notice →US
Siemens Medical Solutions USA, Inc recalls ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material…
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable…
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →US
ORTHO-CLINICAL DIAGNOSTICS recalls VITROS Immunodiagnostic Products Troponin I ES Reagent Pack
Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.
FDA · DeviceORTHO-CLINICAL DIAGNOSTICSWorldwide distribution - US Nationwide and the country of Canada.Official notice →US
Waldemar Link GmbH & Co. KG (Mfg Site) recalls LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 17 - indicated for patients with Mobility-limiting diseases,…
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this…