Drugs & medicalClass III

Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…

Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…

Read the official notice →
Date
2023-07-05
Country
United States
Brand
Randox Laboratories
Product
Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…
Classification
Class III
Units
CA, IL, VA, WA, WI including PR
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Randox Laboratories recalls

  1. 2023-08-30
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…

    As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…

    FDA · DeviceRandox Laboratories Ltd.CA, FL, IL, ME, MI, NC, OH, PA,NJ,Official notice →
  3. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…

    (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…

    FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →
  4. 2023-08-16
  5. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)

    Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…Official notice →
  6. 2023-04-05
  7. US
    Drugs & medicalClass II

    Randox Laboratories recalls CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

    The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

    FDA · DeviceRandox LaboratoriesPA, OH, WVOfficial notice →
  8. 2023-01-11
  9. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical chemistry and…

    Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.Official notice →
  10. 2022-12-14
  11. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834

    Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide Distribution: CA, NC, NJ and WV,Official notice →
All 27 Randox Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2023-07-05
  2. US
    Drugs & medicalClass II

    Hologic, Inc recalls Sertera 14 Gauge Biopsy Device-intended to obtain percutaneous core biopsy samples from soft tissue and tumors of the breast REF: SERTERA-14

    The inner needle of the Sertera device is either detaching or discharging and may cause delay in surgery

    FDA · DeviceHologic, IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    DRE Medical Group Inc recalls Philips Respironics V60 Ventilator-Intended to mechanically control or assist patient breathing by delivering a predetermined percentage…

    Devices were imported with Japanese Language and NIST Oxygen Gas Fittings and retrofitted, may pose a serious health risk as the gas pathway of these devices may be contaminated.

    FDA · DeviceDRE Medical Group IncUS: CA, FLOfficial notice →
  4. US
    Drugs & medicalClass II

    Luminex Corporation recalls VERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-022

    It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result…

    FDA · DeviceLuminex CorporationWorldwide and US Nationwide DistributionOfficial notice →
  5. US
    Drugs & medicalClass II

    Tornier, Inc recalls AEQUALIS FLEX REVIVE Assembly Screw 0mm Short, Model Number ARS655118

    A comingle/swap resulted in the labeling and laser marking on the product to state it is a Aequalis Flex Revive Assembly Screw 0mm Standard (ARS655101) but it was a Aequalis Flex Revive Assembly Screw 0mm Short…

    FDA · DeviceTornier, IncWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →
  6. US
    Drugs & medicalClass II

    Limacorporate S.p.A recalls REF 1975.14.800, Prima TT Genoid Modular Reverse TT Baseplate, Sterile R

    Due to manufacturing issue, their is a potential for peripheral holes to be out of specification.

    FDA · DeviceLimacorporate S.p.AUS Nationwide distribution in the states of MI, IL, ND, MO, OK.Official notice →
  7. US
    Drugs & medicalClass II

    Covidien, LLC recalls Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

    Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.

    FDA · DeviceCovidien, LLCInternational distribution in the country of Taiwan.Official notice →