Drugs & medicalClass II

Randox Laboratories Ltd. recalls IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to…

Read the official notice →
Date
2022-12-14
Country
United States
Brand
Randox Laboratories
Product
IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834
Classification
Class II
Units
US Nationwide Distribution: CA, NC, NJ and WV,
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Randox Laboratories recalls

  1. 2023-07-05
  2. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…

    Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…

    FDA · DeviceRandox Laboratories Ltd.CA, IL, VA, WA, WI including PROfficial notice →
  3. 2023-04-05
  4. US
    Drugs & medicalClass II

    Randox Laboratories recalls CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

    The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

    FDA · DeviceRandox LaboratoriesPA, OH, WVOfficial notice →
  5. 2023-01-11
  6. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical chemistry and…

    Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.Official notice →
  7. 2022-09-28
  8. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV,…

    ***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of WV, NC.Official notice →
  9. 2022-07-27
  10. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0

    Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal…

    FDA · DeviceRandox Laboratories Ltd.Nationwide Distribution - MD, OR, PR, MT, WVOfficial notice →
  11. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD

    A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the state of IL.Official notice →
All 27 Randox Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2022-12-14
  2. US
    Drugs & medicalClass II

    Medacta Usa Inc recalls NextAR Shoulder Surgical Technique, Shoulder NextAR RSA Platform, CT based computer-assisted surgical navigation platform used to perform…

    A warning has been added to the Surgical Technique for NextAR Shoulder, advising verification of suitable patient conditions prior to placement of the NextAR Shoulder Target Holder: WARNING Target fixation requires…

    FDA · DeviceMedacta Usa IncAL, FL, NH, CO, IL, SC, AROfficial notice →
  3. US
    Drugs & medicalClass II

    Bunnell, Inc. recalls 2.5 mm LifePort Endotracheal Tube Adapters 10-pack, Catalog: 9025; and 2.5 mm LifePort Endotracheal Tube Adaptors inside Patient Circuit…

    Some adaptors, for use with high frequency ventilators, were packaged improperly, and may deform at the tip, which may impair the ability to pass a suction catheter or may affect normal ventilator performance, which…

    FDA · DeviceBunnell, Inc.US Nationwide Distribution: PA, NY, AR, AZ, TN, SC, MA, FL, CA, TX, MD, VA,…Official notice →
  4. US
    Drugs & medicalClass II

    Full Vision Inc recalls FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 220V, Model #317-07929

    This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet…

    FDA · DeviceFull Vision IncWi, GA, NY, MexicoOfficial notice →
  5. US
    Drugs & medicalClass II

    Cardinal Health recalls Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 ml) 5ml Catheter Product Code: 2480 Intended for urinary drainage from the…

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of…

    FDA · DeviceCardinal HealthNationwideOfficial notice →
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers…

    Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide Distribution: US (nationwide) to states of: AL, CA, FL, GA, IN, KY,…Official notice →
  7. US
    Drugs & medicalClass I

    Caire, Inc. recalls CAIRE Liberator 45, MODEL 13262253, Liquid Oxygen System Unit

    An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.

    FDA · DeviceCaire, Inc.US Nationwide - Worldwide Distribution: CA, NC, OH, AZ, and Canada, Chile,…Official notice →
See all drugs & medical recalls →