Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers…

Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE…

Read the official notice →
Date
2022-12-14
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers for the…
Classification
Class II
Units
Worldwide Distribution: US (nationwide) to states of: AL, CA, FL, GA, IN, KY,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2023-03-01
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection…

    Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide distribution in the states of GA, UT.Official notice →
  3. 2023-01-18
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/…

    Potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿…Official notice →
  5. 2022-12-21
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Systems 3gAllergy Specific IgE Universal Kit Tests Japan Siemens Material Number (SMN): 10711939 For the…

    Potential for falsely elevated specific IgE mold allergen reactivity with quality control material and patient samples, may result in false reactive result

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution. US Nationwide, Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿…Official notice →
  7. 2022-10-26
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Lithium (Li), Siemens Material Number 11097535. in vitro diagnostic assay

    Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →
  9. US
    Drugs & medicalClass II

    SIEMENS HEALTHCARE DIAGNOSTICS INC recalls N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297

    N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

    FDA · DeviceSIEMENS HEALTHCARE DIAGNOSTICS INCNationwideOfficial notice →
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN):…

    Dimension Vista¿ High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: CanadaOfficial notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2022-12-14
  2. US
    Drugs & medicalClass II

    Full Vision Inc recalls FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 220V, Model #317-07929

    This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet…

    FDA · DeviceFull Vision IncWi, GA, NY, MexicoOfficial notice →
  3. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834

    Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide Distribution: CA, NC, NJ and WV,Official notice →
  4. US
    Drugs & medicalClass II

    Cardinal Health recalls Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 ml) 5ml Catheter Product Code: 2480 Intended for urinary drainage from the…

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of…

    FDA · DeviceCardinal HealthNationwideOfficial notice →
  5. US
    Drugs & medicalClass I

    Caire, Inc. recalls CAIRE Liberator 45, MODEL 13262253, Liquid Oxygen System Unit

    An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.

    FDA · DeviceCaire, Inc.US Nationwide - Worldwide Distribution: CA, NC, OH, AZ, and Canada, Chile,…Official notice →
  6. US
    Drugs & medicalClass I

    Remel Inc recalls Thermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD AST

    An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp.,…

    FDA · DeviceRemel IncUS Nationwide DistributionOfficial notice →
  7. US
    Drugs & medicalClass I

    DEWEI MEDICAL EQUIPMENT CO LTD recalls Dewei DNA/RNA Preservation Kit, Catalog number DW-80005-1B

    Products were distributed in the United States without FDA authorization or clearance.

    FDA · DeviceDEWEI MEDICAL EQUIPMENT CO LTDUS Distribution in states to: CA, FL, KY, and MI.Official notice →
See all drugs & medical recalls →