Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection…

Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2…

Read the official notice →
Date
2023-03-01
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection of antibodies…
Classification
Class II
Units
US Nationwide distribution in the states of GA, UT.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2023-03-22
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus…

    Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile…Official notice →
  3. 2023-03-15
  4. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls epoc Host 2 as follows: Product SMN epoc Host 10736387 epoc Host 2, Canada 10736388 epoc Host 2, China 10736389 epoc Host 2,…

    pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

    FDA · DeviceSiemens Healthcare Diagnostics IncNationwide Foreign: Algeria Argentina Australia Austria Bahrain Belarus…Official notice →
  5. 2023-03-01
  6. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and…

    Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿…Official notice →
  7. 2023-01-18
  8. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/…

    Potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿…Official notice →
  9. 2022-12-21
  10. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Systems 3gAllergy Specific IgE Universal Kit Tests Japan Siemens Material Number (SMN): 10711939 For the…

    Potential for falsely elevated specific IgE mold allergen reactivity with quality control material and patient samples, may result in false reactive result

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution. US Nationwide, Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿…Official notice →
  11. 2022-12-14
  12. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers…

    Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide Distribution: US (nationwide) to states of: AL, CA, FL, GA, IN, KY,…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2023-03-01
  2. US
    Drugs & medicalClass II

    Access Vascular, Inc recalls HydroMID 4Fr Single Lumen Maximal Barrier Kit- Indicated for short term access (<30days) to the peripheral venous access system for…

    Product Mislabeled on the outer bag and inner kit Tyvek header bag

    FDA · DeviceAccess Vascular, IncUS Nationwide distribution in the states of FL, TX, WA.Official notice →
  3. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and…

    Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿…Official notice →
  4. US
    Drugs & medicalClass II

    Hologic, Inc. recalls Hologic 3Dimensions/Selenia Dimensions Models: RM-SDM-00001-3D SDM-00001-2D SDM-00001-3D SDM-05000-3D3 SDM-05000-2DC SDM-00001-M3D…

    C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual

    FDA · DeviceHologic, Inc.US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass I

    Draeger Medical, Inc. recalls NovaStar TS, Standard elbow, Reusable, Material Numbers MP01579 (Size S), MP01580 (Size M), and MP01581 (Size L)

    The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.

    FDA · DeviceDraeger Medical, Inc.US Nationwide DistributionOfficial notice →
  6. US
    Drugs & medicalClass II

    RISING PHARMACEUTICALS recalls Warfarin Sodium Tablets, USP 1 mg, 100-count bottle, Rx Only, Distributed by: Rising Health, LLC, Saddle Brook, NJ 07663, NDC# 57237-119-01

    Failed Impurities/Degradation Specifications

    FDA · DrugRISING PHARMACEUTICALSProduct was distributed to 16 distributors and 8 wholesalers who may have…Official notice →
  7. US
    Drugs & medicalClass II

    B. Braun Medical Inc. recalls 0.9% Sodium Chloride Injection, USP, 500 mL Excel Plus Container, Rx Only, B. Braun Medical Inc. Bethlehem, PA 18018, NDC 0264-5802-10

    Lack of sterility assurance: Recall of certain batches of 0.9% Sodium Chloride for Injection USP in EXCEL¿ Plus IV Container product due to the possibility of an incomplete seal that may cause leakage. The impacted…

    FDA · DrugB. Braun Medical Inc.USA Nationwide.Official notice →