Siemens Healthcare Diagnostics, Inc. recalls Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus…
Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of…
Read the official notice →- Date
- 2023-03-22
- Country
- United States
- Brand
- Siemens Healthcare Diagnostics
- Product
- Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus in human whole…
- Classification
- Class II
- Units
- Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Healthcare Diagnostics recalls
- 2023-07-12
US
Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Storage and Retrieval Module (SRM)
A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide distribution in the states of CA, CO,…Official notice →- 2023-05-31
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use…
Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias
FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide and Worldwide Distribution Foreign: Argentina Australia…Official notice →- 2023-05-10
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number…
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →- 2023-03-15
US
Siemens Healthcare Diagnostics Inc recalls epoc Host 2 as follows: Product SMN epoc Host 10736387 epoc Host 2, Canada 10736388 epoc Host 2, China 10736389 epoc Host 2,…
pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure
FDA · DeviceSiemens Healthcare Diagnostics IncNationwide Foreign: Algeria Argentina Australia Austria Bahrain Belarus…Official notice →- 2023-03-01
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and…
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur XP and ADVIA Centaur XPT Anti-HBe2 (aHBe2) 50T- In vitro diagnostic immunoassay for the qualitative detection…
Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide distribution in the states of GA, UT.Official notice →
Other drugs & medical recalls in the United States
- 2023-03-22
US
Limacorporate S.p.A recalls TEMA Axle Small, REF 1590.15.010 (additional information is pending receipt of the firm's product label/labeling)
Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.
US
Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK
The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if…
FDA · DeviceBard Peripheral Vascular IncWorldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO,…Official notice →US
Brainlab AG recalls Brainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System, model 20910-01E and 20910-01F.
Deep Inspiration Breath Hold (DIBH) functionality may not work as specified when based on a standard-workflow Treatment Template
FDA · DeviceBrainlab AGUS Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA. IL, IN,…Official notice →US
Beckman Coulter, Inc. recalls Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access…
An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay…
FDA · DeviceBeckman Coulter, Inc.Worldwide and Nationwide Distribution. There was government/military…Official notice →US
Stryker Medical Division of Stryker Corporation recalls Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency…
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement…
FDA · DeviceStryker Medical Division of Stryker CorporationUS Nationwide distribution in the states of NY, VA.Official notice →US
Fenwal Inc recalls Amicus Exchange Kit Therapeutics. Component of Amicus Separator multiprocedural apheresis platform
Potential for centrifuge packs to develop a stress leak for certain lots of Amicus MNC Apheresis kit and Amicus Exchange kit (Therapeutic Kits) on the Amicus Separator.
FDA · DeviceFenwal IncNationwide distribution. International distribution to France, Korea, Taiwan,…Official notice →