Drugs & medicalClass II

Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK

The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if…

Read the official notice →
Date
2023-03-22
Country
United States
Brand
Bard Peripheral Vascular
Product
Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK
Classification
Class II
Units
Worldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Bard Peripheral Vascular recalls

  1. 2024-01-17
  2. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM;…

    The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy…

    FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution including in the states of PR, IL, MA,…Official notice →
  3. 2023-10-11
  4. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016

    Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly…

    FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA,…Official notice →
  5. 2023-05-17
  6. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM

    Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and…

    FDA · DeviceBard Peripheral Vascular IncUS Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA,…Official notice →
  7. 2023-02-15
  8. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls The Covera Vascular Covered Stent is a flexible, self-expanding endoprothesis comprised of ePTFE encapsulating a nitinol stent…

    An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.

    FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
  9. 2023-02-08
  10. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Catalog Number / Product Description 0601600 2.7F Broviac CV Catheter Repair Kit 0601610 4.2 Fr Broviac CV Catheter Repair Kit…

    Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).

    FDA · DeviceBard Peripheral Vascular IncWorldwide Distribution: U.S.(nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →
  11. 2022-06-29
  12. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls REF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO,

    There is the potential that a 6 Fr Catheter Drainage kit contains an 8 Fr needle instead of a 6 Fr needle. They can be used in conjunction with the following drainage methods: Universal extension set and drainage bag,…

    FDA · DeviceBard Peripheral Vascular IncWorldwide distribution - US Nationwide distribution in the states of MS, NC,…Official notice →
All 20 Bard Peripheral Vascular recalls →

Other drugs & medical recalls in the United States

  1. 2023-03-29
  2. US
    Drugs & medicalClass II

    Boehringer Ingelheim Pharmaceuticals, Inc. recalls JARDIANCE (Empagliflozin), 25 mg Tablets, packaged in a) 30-count (NDC0597-0153-30) and b) 90-count (NDC…

    Labeling: Label Mix-up

    FDA · DrugBoehringer Ingelheim Pharmaceuticals, Inc.Product was distributed nationwide within the United States and PROfficial notice →
  3. 2023-03-22
  4. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls Patient Side Cart, REF: 380620, with instrument arms, used as part of the da Vinci X Surgical System, REF: IS4200; Patient Side Cart,…

    Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled…

    FDA · DeviceIntuitive Surgical, Inc.Worldwide - US Nationwide distribution in the states of FL, GA, WA and the…Official notice →
  5. US
    Drugs & medicalClass II

    Limacorporate S.p.A recalls TEMA Axle Small, REF 1590.15.010 (additional information is pending receipt of the firm's product label/labeling)

    Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.

    FDA · DeviceLimacorporate S.p.AU.S. Distribution to: TXOfficial notice →
  6. US
    Drugs & medicalClass II

    Brainlab AG recalls Brainlab ExacTrac Dynamic software, Medical Charged-Particle Radiation Therapy System, model 20910-01E and 20910-01F.

    Deep Inspiration Breath Hold (DIBH) functionality may not work as specified when based on a standard-workflow Treatment Template

    FDA · DeviceBrainlab AGUS Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA. IL, IN,…Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus…

    Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile…Official notice →
  8. US
    Drugs & medicalClass II

    Beckman Coulter, Inc. recalls Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access…

    An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay…

    FDA · DeviceBeckman Coulter, Inc.Worldwide and Nationwide Distribution. There was government/military…Official notice →