Drugs & medicalClass II

Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016

Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly…

Read the official notice →
Date
2023-10-11
Country
United States
Brand
Bard Peripheral Vascular
Product
Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Bard Peripheral Vascular recalls

  1. 2024-09-18
  2. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen…

    Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur…

    FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The…Official notice →
  3. 2024-01-17
  4. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls BD Recanalization System, REF: BDRECANSYSTEM

    Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute…

    FDA · DeviceBard Peripheral Vascular IncUS: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS,…Official notice →
  5. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM;…

    The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy…

    FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution including in the states of PR, IL, MA,…Official notice →
  6. 2023-05-17
  7. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit: REF: 2010MSK, 20G x 10CM; REF: 2016MSK, 20G x 16CM

    Coaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and…

    FDA · DeviceBard Peripheral Vascular IncUS Nationwide Distribution: PR, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA,…Official notice →
  8. 2023-03-22
  9. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit: 16G x 10CM, REF: 1610MSK; 18G x 16CM, REF: 1816MSK

    The external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if…

    FDA · DeviceBard Peripheral Vascular IncWorldwide Distribution: US (Nationwide) includes states of: FL, MD, OH, CO,…Official notice →
  10. 2023-02-15
  11. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls The Covera Vascular Covered Stent is a flexible, self-expanding endoprothesis comprised of ePTFE encapsulating a nitinol stent…

    An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.

    FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →
All 20 Bard Peripheral Vascular recalls →

Other drugs & medical recalls in the United States

  1. 2023-10-11
  2. US
    Drugs & medicalClass II

    Blue Belt Technologies, Inc recalls Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

    CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational…

    FDA · DeviceBlue Belt Technologies, IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    AB Sciex, LLC recalls Citrine Triple Quad Mass Spectrometer, Part Number 5063684

    In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly…

    FDA · DeviceAB Sciex, LLCUS Nationwide.Official notice →
  4. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as…

    Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive)…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Becton Dickinson Infusion Therapy Systems Inc. recalls REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min ,…

    There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point…

    FDA · DeviceBecton Dickinson Infusion Therapy Systems Inc.US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ,…Official notice →
  6. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls Guider Softip Guide Catheter 5F 100CM MULTI PURPOSE, Catalog number M003101630, cardiac guide catheter

    Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from…

    FDA · DeviceBoston Scientific CorporationUS Nationwide distribution in the state of Indiana.Official notice →
  7. US
    Drugs & medicalClass II

    Ortho-Clinical Diagnostics, Inc. recalls Proboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry…

    Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an…

    FDA · DeviceOrtho-Clinical Diagnostics, Inc.Worldwide - US Nationwide distribution.Official notice →
See all drugs & medical recalls →