Bard Peripheral Vascular Inc recalls Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen…
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur…
Read the official notice →- Date
- 2024-09-18
- Country
- United States
- Brand
- Bard Peripheral Vascular
- Product
- Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis…
- Classification
- Class II
- Units
- Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Bard Peripheral Vascular recalls
- 2024-11-27
US
Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF:…
Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle…
FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC,…Official notice →- 2024-11-20
US
Bard Peripheral Vascular Inc recalls The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a…
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead…
FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →- 2024-10-02
US
Bard Peripheral Vascular Inc recalls BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy System…
Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised.
FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution in the Puerto Rico and the countries…Official notice →- 2024-01-17
US
Bard Peripheral Vascular Inc recalls BD Recanalization System, REF: BDRECANSYSTEM
Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute…
FDA · DeviceBard Peripheral Vascular IncUS: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS,…Official notice →US
Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM;…
The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy…
FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution including in the states of PR, IL, MA,…Official notice →- 2023-10-11
US
Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016
Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly…
FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA,…Official notice →
Other drugs & medical recalls in the United States
- 2024-09-18
US
Ultradent Products, Inc. recalls Brand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Number: 4757- Software Version: N/A…
Due to a manufacturing issue, red particulates present in the expressed composite.
FDA · DeviceUltradent Products, Inc.US: AZ, CA,CO,FL,GA,HI, IA, ID,IL,IN,KS,KY,LA, MA, MD, MI, MN, MO, MS, NC, ND,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to…
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Domestic: Nationwide Distribution. International: Austria, Belgium, Canada,…Official notice →US
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11097543
The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, GA, IA, IL,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls Medline Convenience kits, labeled as: 1) SUTURING SET PACK ASSY P791011, Pack Number 59112; 2) SUTURE SET, Pack Number…
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
US
PACIFIC LASERTECH, LLC recalls 25-LHP-928, HeNe Laser System, 25 mW
These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall…
US
Smiths Medical ASD Inc. recalls Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NXX
There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.