Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF:…
Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle…
Read the official notice →- Date
- 2024-11-27
- Country
- United States
- Brand
- Bard Peripheral Vascular
- Product
- Bard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEE 12G X 10 CM, REF: MQ1210;…
- Classification
- Class II
- Units
- Worldwide - US Nationwide including in the states of WA, LA, MN, PA, UT, SC,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Bard Peripheral Vascular recalls
- 2025-12-17
US
Bard Peripheral Vascular Inc recalls Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt…
FDA · DeviceBard Peripheral Vascular IncUS Nationwide distribution in the states of CA, MN, IN, MI, LA.Official notice →- 2025-10-08
US
Bard Peripheral Vascular Inc recalls Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF…
Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use.…
FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution including in the states of NJ, NY, FL,…Official notice →- 2025-03-12
US
Bard Peripheral Vascular Inc recalls Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F…
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could…
FDA · DeviceBard Peripheral Vascular IncUS Nationwide Distribution: NY, VA, KY, IN, OR, FL, MD, NJ, LA, CO, PR, PA,…Official notice →- 2024-11-20
US
Bard Peripheral Vascular Inc recalls The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a…
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead…
FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution including in the states of AK, AL,…Official notice →- 2024-10-02
US
Bard Peripheral Vascular Inc recalls BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy System…
Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised.
FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution in the Puerto Rico and the countries…Official notice →- 2024-09-18
US
Bard Peripheral Vascular Inc recalls Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen…
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur…
FDA · DeviceBard Peripheral Vascular IncWorldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The…Official notice →
Other drugs & medical recalls in the United States
- 2024-11-27
US
Caire, Inc. recalls CAIRE FreeStyle Comfort portable oxygen concentrator, model #AS200-2.
Potential failure of affected units to recharge the battery of the oxygen concentrators.
FDA · DeviceCaire, Inc.Worldwide distribution - US Nationwide, including PR and the countries of…Official notice →US
Biomet, Inc. recalls JuggerLoc/ToggleLoc 2.9mm Drill Guide, Model number 110027357
Firm received complaints reporting that guides were warped when pulled out of packaging. This issue may result in extension of surgery to find another product, or may lead to adverse local tissue reaction leading to…
FDA · DeviceBiomet, Inc.Worldwide distribution - US Nationwide and the countries of Australia, Brazil,…Official notice →US
Hobbs Medical, Inc. recalls Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the…
FDA · DeviceHobbs Medical, Inc.Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN,…Official notice →US
Smiths Medical ASD, Inc. recalls smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis…
The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.
US
Baxter Healthcare Corporation recalls Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4482; use in Peritoneal Dialysis
Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using…
FDA · DeviceBaxter Healthcare CorporationWorldwide Distribution (US Nationwide distribution)Official notice →US
Evaric Pharmaceuticals Inc. recalls Lisinopril Tablets, USP 10 mg, 90 tablets per bottle, Rx Only, Distributed by: Walmart, Bentonville, AR 72716, Manufactured for:…
Presence of Foreign Object: A pharmacist discovered a metal fragment embedded in a lisinopril 10 mg tablet.