Drugs & medicalClass II

AB Sciex, LLC recalls Citrine Triple Quad Mass Spectrometer, Part Number 5063684

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly…

Read the official notice →
Date
2023-10-11
Country
United States
Brand
AB Sciex
Product
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Classification
Class II
Units
US Nationwide.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More AB Sciex recalls

  1. 2021-11-03
  2. US
    Drugs & medicalClass II

    AB SCIEX recalls Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem…

    The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making…

    FDA · DeviceAB SCIEXWorldwide distribution - US Nationwide distribution in the states of District…Official notice →
All 2 AB Sciex recalls →

Other drugs & medical recalls in the United States

  1. 2023-10-11
  2. US
    Drugs & medicalClass II

    ROi CPS LLC recalls GS0008 2Y - Burn Pack - STL, Item Number 880082025

    Medical convenience kit contains Zimmer Dermatome Blades which were recalled by Zimmer due to 38 complaints received related to skin grafts being thin and non-uniform when using the affected blades.

    FDA · DeviceROi CPS LLCMissouriOfficial notice →
  3. US
    Drugs & medicalClass II

    Zimmer Surgical Inc recalls Zimmer Dermatome Blades- 10-pack boxes of blades. Intended for use with the dermatome handpiece which is a skin grafting instrument that…

    Skin grafts thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area

    FDA · DeviceZimmer Surgical IncUS Nationwide Foreign: AUSTRALIA BRAZIL CHINA HONG KONG INDIA JAPAN MALAYSIA…Official notice →
  4. US
    Drugs & medicalClass II

    Blue Belt Technologies, Inc recalls Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

    CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational…

    FDA · DeviceBlue Belt Technologies, IncUS Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as…

    Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive)…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Bard Peripheral Vascular Inc recalls Bard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016

    Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly…

    FDA · DeviceBard Peripheral Vascular IncWorldwide - US Nationwide distribution in the states of PR, CO, TX, CA, WA,…Official notice →
  7. US
    Drugs & medicalClass II

    Becton Dickinson Infusion Therapy Systems Inc. recalls REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min ,…

    There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point…

    FDA · DeviceBecton Dickinson Infusion Therapy Systems Inc.US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ,…Official notice →
See all drugs & medical recalls →