- 2023-10-11
US
AB Sciex, LLC recalls Citrine Triple Quad Mass Spectrometer, Part Number 5063684
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly…
- 2021-11-03
US
AB SCIEX recalls Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem…
The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making…