Drugs & medicalClass II

Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Storage and Retrieval Module (SRM)

A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the…

Read the official notice →
Date
2023-07-12
Country
United States
Brand
Siemens Healthcare Diagnostics
Product
Aptio Automation Storage and Retrieval Module (SRM)
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of CA, CO,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Healthcare Diagnostics recalls

  1. 2023-09-06
  2. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl…

    Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿CH Total Bilirubin_2 (TBil_2),…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay VietnamOfficial notice →
  3. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human…

    Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Country Albania Algeria Argentina Australia Austria…Official notice →
  4. 2023-08-30
  5. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the…

    There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA,…Official notice →
  6. 2023-05-31
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use…

    Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.US Nationwide and Worldwide Distribution Foreign: Argentina Australia…Official notice →
  8. 2023-05-10
  9. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Cortisol 50T (Material Number 10995538), 250T (Material Number 10995537), and REF 250T (Material Number…

    There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →
  10. 2023-03-22
  11. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus…

    Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of…

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile…Official notice →
All 80 Siemens Healthcare Diagnostics recalls →

Other drugs & medical recalls in the United States

  1. 2023-07-12
  2. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191,…

    Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock.

    FDA · DeviceCareFusion 303, Inc.US Nationwide-Worldwide Distribution: CO, MS, HI, FL, GA, CA, ND, GU, DE, MO,…Official notice →
  3. US
    Drugs & medicalClass III

    NeuMoDx Molecular Inc recalls NeuMoDx SARS-CoV-2 Assay -IVD intended for the qualitative detection of SARS-CoV-2 RNA from nasal, nasopharyngeal and oropharyngeal swabs…

    Potential for lot number discrepancies between the test strip packaging and the barcode label applied to the test strip that is read by the instrument for LOT 123909 of NeuMoDx SARS-CoV-2 Test Strips (REF 300800). The…

    FDA · DeviceNeuMoDx Molecular IncUS Nationwide distribution in the states of FL, MI, NM, NY, TN, TX.Official notice →
  4. US
    Drugs & medicalClass II

    Zimmer, Inc. recalls NexGen Precoat Stemmed Tibial Plate Size 5

    Device outer packaging was incorrectly labeled.

    FDA · DeviceZimmer, Inc.International distribution in the country of India.Official notice →
  5. US
    Drugs & medicalClass II

    DeRoyal Industries Inc recalls Custom medical procedure packs labeled as: a) CATH ANGIOGRAPHY DRAPE PACK, REF 89-7102; b) CHOLANGIOGRAPHY KIT, REF 89-7322; c) CATH LAB…

    The outer package seal could be open, or compromised which affects the sterility of the procedure pack inside.

    FDA · DeviceDeRoyal Industries IncUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    TEI Biosciences, Inc. recalls PriMatrix Ag Meshed; Dermal Repair Scaffold; Rx Only; Sizes 10x12 cm (2:1), 10x25 cm (2:1), 20x25 cm (1:1), 4x4 cm (2:1), 6x6 cm (2:1),…

    Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing.

    FDA · DeviceTEI Biosciences, Inc.Domestic: Nationwide Distribution.Official notice →
  7. US
    Drugs & medicalClass II

    NeuroLogica Corporation recalls OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and…

    1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain…

    FDA · DeviceNeuroLogica CorporationWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →
See all drugs & medical recalls →