Randox Laboratories recalls CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826
The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range
Read the official notice →- Date
- 2023-04-05
- Country
- United States
- Brand
- Randox Laboratories
- Product
- CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826
- Classification
- Class II
- Units
- PA, OH, WV
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Randox Laboratories recalls
- 2023-08-30
US
Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…
FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →- 2023-08-16
US
Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…Official notice →- 2023-07-05
US
Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…
- 2023-01-11
US
Randox Laboratories Ltd. recalls Liquid Assayed Specific Protein Control Level 2- In vitro diagnostic use, in the quality control of serum on clinical chemistry and…
Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.Official notice →- 2022-12-14
US
Randox Laboratories Ltd. recalls IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834
Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to…
- 2022-09-28
US
Randox Laboratories Ltd. recalls The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV,…
***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay…
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of WV, NC.Official notice →
Other drugs & medical recalls in the United States
- 2023-04-05
US
Brainlab AG recalls Cirq Arm System 2.0
Risk of mechanical instability due to potential manufacturing error of the device
US
Alcon Research, LLC recalls Alcon Custom Pak Surgical Procedure Pack: Pak Number & Pak Description: 17533-04 DR GROSS LPS LLC 18704-03 CATARACT SURGI CARE 18382-03 DR…
Sterility; Steril surgical procedure packs contain incomplete seals.
FDA · DeviceAlcon Research, LLCUS: AL, AR, CA, IA, ID, IN, KY, MN, NC, ND, NJ, OH, PA, TX, OUS: NoneOfficial notice →US
GAGA PRO LIGHTING EQUIPMENT CO., recalls GaGa Pro Lighting Equipment Co. Ltd. manufactures Class IV Laser Light Show (LLS)projectors for the purposes of demonstration…
GAGA Pro LLS systems was not in compliance with the FDA's Performance Standard for laser products.
US
Preat Corp recalls Neodent GM X 6mm Engaging Titanium Base, REF: 9007162
Engaging titanium base contains a dimensional condition that allows vertical movement of the titanium base after the prosthetic screw is torqued to the recommended value, which if undetected and installed on the…
US
Ascend Laboratories, LLC recalls Dabigatran Etexilate 150mg Capsules, 150 mg, Rx Only, 60 Capsules per bottle, Manufactured by: Alkem Laboratories Ltd., INDIA,…
CGMP Deviations: Detection of N-nitroso-dabigatran (NDAB) impurity levels above the Acceptable Daily Intake Limit.
US
Apollo Care recalls Fentanyl 500mcg (2mcg/mL) and Ropivacaine HCl 250mg (1mg/mL) added to 250mL 0.9% Sodium Chloride, For Epidural Use Only, Rx only, 3801…
Lack of assurance of sterility: Suspected microbial growth present on external label packaging.