Drugs & medicalClass II

Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…

(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…

Read the official notice →
Date
2023-08-30
Country
United States
Brand
Randox Laboratories
Product
Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol,…
Classification
Class II
Units
Worldwide distribution - US Nationwide distribution in the states of AR, FL,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Randox Laboratories recalls

  1. 2023-11-08
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000

    RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution including Puerto RicoOfficial notice →
  3. 2023-09-27
  4. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215

    there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet…

    FDA · DeviceRandox Laboratories Ltd.US States: GA, OH, OR, PA, TX, VAOfficial notice →
  5. 2023-08-30
  6. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…

    As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…

    FDA · DeviceRandox Laboratories Ltd.CA, FL, IL, ME, MI, NC, OH, PA,NJ,Official notice →
  7. 2023-08-16
  8. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)

    Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…Official notice →
  9. 2023-07-05
  10. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…

    Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…

    FDA · DeviceRandox Laboratories Ltd.CA, IL, VA, WA, WI including PROfficial notice →
  11. 2023-04-05
  12. US
    Drugs & medicalClass II

    Randox Laboratories recalls CRP Immunoturbidimetric reagent. Intended for the quantitative in vitro determination of CRP concentration in serum Catalogue number: CP3826

    The Antibody (R2 reagent) is showing a positive bias compared to previous batches on patient samples. Quality Controls do not show the same bias and will be within range

    FDA · DeviceRandox LaboratoriesPA, OH, WVOfficial notice →
All 27 Randox Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-30
  2. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc.-Littleton recalls The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric…

    Ground cable installed incorrectly.

    FDA · DeviceMedtronic Navigation, Inc.-LittletonWorldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL,…Official notice →
  3. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…

    As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…

    FDA · DeviceRandox Laboratories Ltd.CA, FL, IL, ME, MI, NC, OH, PA,NJ,Official notice →
  4. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC…

    The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body…

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Invacare Corporation recalls DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654807 - DOLOMITE…

    Premature failure of the seat during use, the plastic eyelets of the folding mechanism of the device's seat can break, and in a worst-case crossbar also breaks and the Rollator can collapse causing injury to user

    FDA · DeviceInvacare CorporationUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail

    A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE…

    FDA · DeviceSmith & Nephew, Inc.US, AE, CA, DE, ES, GB, MX, PA, PLOfficial notice →
  7. US
    Drugs & medicalClass II

    Boston Scientific Corporation recalls EMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator

    There is a potential for a rare interaction between the EMBLEM S-ICD and LATITUDE communicator, which may cause S-ICD sensing disablement for a 24-hour interval.

    FDA · DeviceBoston Scientific CorporationWorldwideOfficial notice →
See all drugs & medical recalls →