Randox Laboratories Ltd. recalls Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215
there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet…
Read the official notice →- Date
- 2023-09-27
- Country
- United States
- Brand
- Randox Laboratories
- Product
- Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215
- Classification
- Class II
- Units
- US States: GA, OH, OR, PA, TX, VA
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Randox Laboratories recalls
- 2024-01-10
US
Randox Laboratories Ltd. recalls SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus…
A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.
- 2023-12-13
US
Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.
Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of AR, GA, NC, TN, & PR.Official notice →- 2023-11-08
US
Randox Laboratories Ltd. recalls RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature…
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution including Puerto RicoOfficial notice →- 2023-08-30
US
Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…
As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…
US
Randox Laboratories Ltd. recalls Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride,…
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in…
FDA · DeviceRandox Laboratories Ltd.Worldwide distribution - US Nationwide distribution in the states of AR, FL,…Official notice →- 2023-08-16
US
Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…Official notice →
Other drugs & medical recalls in the United States
- 2023-09-27
US
Abbott Medical recalls Gallant VR Implantable Cardioverter Defibrillator: VR ICD REF CDVRA500Q Gallant DR Implantable Cardioverter Defibrillator: DR ICD REF…
Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of…
FDA · DeviceAbbott MedicalU.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech…Official notice →US
GE Medical Systems Information Technologies Inc recalls Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX…
A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.
FDA · DeviceGE Medical Systems Information Technologies IncDistribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA,…Official notice →US
Tytek Medical Inc recalls TyTek Tension Pneumothorax Access Kit TPAK 14 gauge 3.25 inch- Intended for introduction into the body to facilitate removal of air from…
Updated IFU to inlclude: (1) identify the risks associated with the hazard of needle decompression and (2) improve the artwork showing placement site location
FDA · DeviceTytek Medical IncOH, SC, VA Foreign: Country Australia Denmark Finland Germany Hong Kong/China…Official notice →US
Philips North America Llc recalls Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number:…
Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central…
FDA · DevicePhilips North America LlcNationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark…Official notice →US
Encore Medical, LP recalls DJO Surgical: Empowr Knee, Universal Tibia Baseplate, 6R, REF: 352-02-106
Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to…
FDA · DeviceEncore Medical, LPUS: AL, AZ, CA, CO, FL, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH,…Official notice →US
Stryker Corporation recalls Stryker ACM (Advanced Cement Mixing) System: ACM Kit w/ Femoral Breakaway Nozzle-Allows the user to mix bone cement witha constant high…
Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and…
FDA · DeviceStryker CorporationNationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy…Official notice →