Drugs & medicalClass II

Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.

Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.

Read the official notice →
Date
2023-12-13
Country
United States
Brand
Randox Laboratories
Product
RANDOX Total Bilirubin.
Classification
Class II
Units
US Nationwide distribution in the states of AR, GA, NC, TN, & PR.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Randox Laboratories recalls

  1. 2024-01-24
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351

    CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

    FDA · DeviceRandox Laboratories Ltd.US Nationwide and Puerto Rico.Official notice →
  3. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)

    There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution.Official notice →
  4. 2024-01-10
  5. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus…

    A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.

    FDA · DeviceRandox Laboratories Ltd.NY, WVOfficial notice →
  6. 2023-11-08
  7. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls RX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000

    RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution including Puerto RicoOfficial notice →
  8. 2023-09-27
  9. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Liquid Assayed Chemistry Control Premium Plus Level 3, Catalog Number LAE4215

    there has been a transcription error in the Instructions for Use (IFU) for the Liquid Assayed Chemistry Control Premium Plus Level 3. For Caeruloplasmin, the one and two standard deviation values printed on the sheet…

    FDA · DeviceRandox Laboratories Ltd.US States: GA, OH, OR, PA, TX, VAOfficial notice →
  10. 2023-08-30
  11. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement…

    As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The…

    FDA · DeviceRandox Laboratories Ltd.CA, FL, IL, ME, MI, NC, OH, PA,NJ,Official notice →
All 27 Randox Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2023-12-13
  2. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls Stay Safe Cap

    Updating the labeling to include additional warnings and contraindications for potential side effects.

    FDA · DeviceFresenius Medical Care Holdings, Inc.US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Liberty Medical Solutions, LLC recalls Multi-Mode Stimulator, Transcutaneous Electrical Nerve Stimulation

    The instruction manuals that were included with the Multi-Mode Stimulators distributed from 2018 to 2023 did not contain all of the appropriate warnings and contraindications.

    FDA · DeviceLiberty Medical Solutions, LLCUS Nationwide distribution in the states of CT, FL, NC, NJ, NY.Official notice →
  4. US
    Drugs & medicalClass II

    CooperSurgical, Inc. recalls Cooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr -For administering contrast media during Hysterosalpingography or Hysterosonography procedures…

    Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure

    FDA · DeviceCooperSurgical, Inc.Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLANDOfficial notice →
  5. US
    Drugs & medicalClass II

    Luminex Corporation recalls Luminex VERIGENE BC-GP Utility Tray (5-Pack Carrier), Gram-Positive Blood Culture Nucleic Acid Test, REF 20-011-018, a component of the…

    Potentially defective utility trays in the reagent kits.

    FDA · DeviceLuminex CorporationWorldwide - US Nationwide distribution in the states of AZ, CA, CO, DE, FL,…Official notice →
  6. US
    Drugs & medicalClass II

    Integra LifeSciences Corp. recalls Integra Cranial Access Kit - For access to the subarachnoid space or the lateral ventricles of the brain. Catalog Number: INS7040

    Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device

    FDA · DeviceIntegra LifeSciences Corp.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  7. US
    Drugs & medicalClass II

    Integra LifeSciences Corp. recalls Integra Cranial access kit (without Prep solutions) Bit and Guard-For access to the subarachnoid space or the lateral ventricles of…

    Defect in the outer packaging of the Cranial Access Kits can cause the packaging to split without any additional forces outside compromising the sterility of the device

    FDA · DeviceIntegra LifeSciences Corp.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
See all drugs & medical recalls →