Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

Read the official notice →
Date
2024-01-24
Country
United States
Brand
Randox Laboratories
Product
Calibration Serum Level 3 CAL2351
Classification
Class II
Units
US Nationwide and Puerto Rico.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Randox Laboratories recalls

  1. 2024-03-27
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Liquid Protein Calibrators. C3 and Haptoglobin liquid protein calibrators used with Randox IgA, IgG, and IgM assays that require sample…

    The firm has realigned C3 and Haptoglobin in Liquid Protein Calibrators, IT2691, to reference material ERM-DA470k/IFCC.

    FDA · DeviceRandox Laboratories Ltd.Domestic: CA, FL, IL, ME, NJ, OH, & OR.Official notice →
  3. 2024-02-14
  4. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls RX Series Copper (Cu) Assay Ref. Number CU2340

    Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of SC and OK.Official notice →
  5. 2024-01-24
  6. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

    Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

    FDA · DeviceRandox Laboratories Ltd.US Nationwide including Puerto Rico.Official notice →
  7. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)

    There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution.Official notice →
  8. 2024-01-10
  9. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus…

    A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.

    FDA · DeviceRandox Laboratories Ltd.NY, WVOfficial notice →
  10. 2023-12-13
  11. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.

    Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of AR, GA, NC, TN, & PR.Official notice →
All 27 Randox Laboratories recalls →

Other drugs & medical recalls in the United States

  1. 2024-01-24
  2. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C,…

    A potential for a light system to fall in the operating room.

    FDA · DeviceGetinge Usa Sales IncUS Nationwide. Global Distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls CardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510

    The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard…

    FDA · DeviceCardinal Health 200, LLCUS, Canada, Australia and New ZealandOfficial notice →
  4. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

    Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

    FDA · DeviceRandox Laboratories Ltd.US Nationwide including Puerto Rico.Official notice →
  5. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Calibration Serum Level 3. CAL 3, REF CAL2351. Human Assayed Multi-Sera Level 2 (HN1530) Human Assayed Multi-Sera Level 3 (HE 1532)

    There have been transcription errors on the Instructions For Use (IFU) of the Calibration Serum Level 3 (CAL2351 lot 1260UE, 1262UE and 1315UE) for target values of Bilirubin and Triglycerides, on the IFU of Human…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Xcision Medical Systems, LLC recalls GammaPod - Treatment Planning System-a component of the GammaPod Irradiation System. Ref: XMSGP030A09.07-0.01 Intended for use in…

    Wrong structures imported into the GammaPod Treatment Planning System.-system allowed an RTSS generated from a previous CT scan to be selected

    FDA · DeviceXcision Medical Systems, LLCWorldwide - US Nationwide distribution in the states of MD, PA, TX and the…Official notice →
  7. US
    Drugs & medicalClass II

    Waldemar Link GmbH & Co. KG (Mfg Site) recalls Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee…

    Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to…

    FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)US Nationwide distribution in the state of NH.Official notice →
See all drugs & medical recalls →