Getinge Usa Sales Inc recalls Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C,…
A potential for a light system to fall in the operating room.
Read the official notice →- Date
- 2024-01-24
- Country
- United States
- Brand
- Getinge Usa Sales
- Product
- Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C,…
- Classification
- Class II
- Units
- US Nationwide. Global Distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Getinge Usa Sales recalls
- 2024-07-17
US
Getinge Usa Sales Inc recalls ALPHAMAXX Mobile Operating Table, Model Number/Part Number: 113322B4
The recall is prompted by a defective charger unit (0970.2873) for models 1133.22B4, 1133.22B5, and 1133.22F5 of the ALPHAMAXX mobile operating table. This issue poses a risk of inadequate charging and operational…
FDA · DeviceGetinge Usa Sales IncDomestic: KY, NC, NY, VA. International: Australia, Bulgaria, Croatia, Czech…Official notice →- 2024-06-12
US
Getinge Usa Sales Inc recalls Vaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of…
FDA · DeviceGetinge Usa Sales IncDomestic: CA, CO, IN, MO, OR, VA, WA. International: Aruba, Australia,…Official notice →- 2024-02-14
US
Getinge Usa Sales Inc recalls Flow-c Anesthesia System, model 6887700
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous…
FDA · DeviceGetinge Usa Sales IncWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO,…Official notice →- 2023-10-11
US
Getinge Usa Sales Inc recalls Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as…
Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive)…
- 2023-02-01
US
Getinge Usa Sales Inc recalls Getinge Flow-e Anesthesia System, Part No. 6887900
Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and…
FDA · DeviceGetinge Usa Sales IncUS nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ,…Official notice →- 2022-12-14
US
Getinge Usa Sales Inc recalls Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)
The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power…
FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.Official notice →
Other drugs & medical recalls in the United States
- 2024-01-31
US
Glenmark Pharmaceuticals Inc., USA recalls Fluocinolone Acetonide Oil, 0.01% (Ear Drops), package in 1 Oz. (20 mL fill volume) bottle, Rx only, Manufactured by: Glenmark…
Failed Excipient Specifications: OOS for assay of Isopropyl Alcohol
US
US
Amerisource Health Services LLC recalls Benzonatate Capsules, USP, 100 mg, 100 Capsules (10 capsules x 10 unit dose cards), Rx only, Distributed by: American Health…
Superpotent drug: Assay results were slightly above specification at the time zero point.
- 2024-01-24
US
Cardinal Health 200, LLC recalls CardinalHealth Suction Canister Liner, MEDI-VAC CRD, REF 65651-510
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard…
US
Randox Laboratories Ltd. recalls Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351
Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples
US
Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351
CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.