Getinge Usa Sales Inc recalls Getinge Flow-e Anesthesia System, Part No. 6887900

Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and…

Read the official notice →
Date
2023-02-01
Country
United States
Brand
Getinge Usa Sales
Product
Getinge Flow-e Anesthesia System, Part No. 6887900
Classification
Class I
Units
US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Getinge Usa Sales recalls

  1. 2024-02-14
  2. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Flow-c Anesthesia System, model 6887700

    Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous…

    FDA · DeviceGetinge Usa Sales IncWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO,…Official notice →
  3. 2024-01-24
  4. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Maquet PowerLED/HLED and PowerLED300 OR Light Systems, Model Numbers and Catalog Numbers EPS MB1 - ARDINT059002A, EPS MB1 - ARDINT059002C,…

    A potential for a light system to fall in the operating room.

    FDA · DeviceGetinge Usa Sales IncUS Nationwide. Global Distribution.Official notice →
  5. 2023-10-11
  6. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as…

    Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive)…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution.Official notice →
  7. 2022-12-14
  8. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

    The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.Official notice →
  9. 2022-11-23
  10. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN

    The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.

    FDA · DeviceGetinge Usa Sales IncWorldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN.…Official notice →
  11. 2022-08-31
  12. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Servo-air Ventilator System-The Servo-air Ventilator System is: Intended for respiratory support, monitoring and treatment of pediatric…

    Device does not meet regulatory requirements of stability while stationary, which requires mechanical equipment, other than fixed mechanical equipment, that is intended to be used on the floor or on a table shall be…

    FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide and the countries of Australia (AU),…Official notice →
All 17 Getinge Usa Sales recalls →

Other drugs & medical recalls in the United States

  1. 2023-02-01
  2. US
    Drugs & medicalClass II

    Johnson & Johnson Surgical Vision, Inc. recalls FUSION Dual Pump Packs, REF: OPO73

    Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and…

    FDA · DeviceJohnson & Johnson Surgical Vision, Inc.US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France,…Official notice →
  3. US
    Drugs & medicalClass II

    Agfa Healthcare NV recalls Agfa HealthCare Enterprise Imaging XERO Viewer

    There is a software defect that can cause issues with images.

    FDA · DeviceAgfa Healthcare NVDistribution throughout US OUS distribution to Australia, Belgium, Brazil,…Official notice →
  4. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the…

    In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped

    FDA · DeviceSiemens Medical Solutions USA, IncNationwideOfficial notice →
  5. US
    Drugs & medicalClass II

    Pharmacy Innovations recalls GLUTATHIONE L REDUCED 100MG/ML (2ML NEBVL) INHALATION; GLUTATHIONE L REDUCED 200MG/ML (1ML NEBVL) INHALATION; GLUTATHIONE L REDUCED…

    Lack of Assurance of Sterility: FDA inspection revealed insanitary conditions at the facility.

    FDA · DrugPharmacy InnovationsNationwide in the USAOfficial notice →
  6. US
    Drugs & medicalClass I

    Pfizer Inc. recalls Vancomycin Hydrochloride for Injection, USP, 1.5 g/vial, Sterile powder, Single-dose Fliptop Vial, 10 vials per carton, Rx only,…

    Presence of Particulate Matter: Glass particulate matter detected in injectable.

    FDA · DrugPfizer Inc.USA nationwide and Puerto RicoOfficial notice →
  7. US
    Drugs & medicalClass I

    Glaxosmithkline Consumer Healthcare Holdings DBA Haleon recalls Advil Liqui Gels (minis), Solubilized ibuprofen capsules, 200mg, Pain Reliever Fever Reducer (NSAID),…

    Labeling: Illegible Label; the adhesive migrated onto a portion of the label, causing it to tear when peeled back

    FDA · DrugGlaxosmithkline Consumer Healthcare Holdings DBA HaleonNationwide in the USA.Official notice →
See all drugs & medical recalls →