Siemens Medical Solutions USA, Inc recalls Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the…
In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped
Read the official notice →- Date
- 2023-02-01
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT…
- Classification
- Class II
- Units
- Nationwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2023-03-01
US
Siemens Medical Solutions USA, Inc recalls Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model…
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- 2023-02-22
US
Siemens Medical Solutions USA, Inc. recalls SIEMENS Biograph mCT: a) Flow 64-4R Seismic, b) S(40)-3R Seismic, c) S(64)-3R Seismic, d) S(40)-4R Seismic, e) S(64)-4R…
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - US Nationwide and the countries of Japan, Chile,…Official notice →- 2023-02-08
US
Siemens Medical Solutions USA, Inc recalls System Material # Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee multi-purpose 10094139 Artis zee biplane…
In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue…
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution.Official notice →- 2023-01-25
US
Siemens Medical Solutions USA, Inc recalls Luminos Agile Max, Luminos dRF Max, and LUMINOS Lotus Max
There is potential for the footrest to detach from the patient table during use. Device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system has medical…
US
Siemens Medical Solutions USA, Inc recalls ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor
Mechanical connection between the tabletop and table base may be lost.
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of Armenia,…Official notice →- 2023-01-11
US
Siemens Medical Solutions USA, Inc recalls Programmable Diagnostic Computer
The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during…
FDA · DeviceSiemens Medical Solutions USA, IncDomestic distribution nationwide. Worldwide foreign distribution to Albania…Official notice →
Other drugs & medical recalls in the United States
- 2023-02-01
US
Meridian Bioscience Inc recalls Revogene SARS-CoV-2_IVD real-time RT-PCR test intended for the qualitative detection of RNA from SARSCoV-2 Catalog Number: 410700
Leakage of the original specimen from the plastic microfluidic cartridge (PIE), possibility that laboratory personnel may be exposed to aerosolized specimen containing SARS-CoV-2
FDA · DeviceMeridian Bioscience IncUS Nationwide distribution in the states of AZ, CA, IN, KS, KY, LA, MI, MN,…Official notice →US
Johnson & Johnson Surgical Vision, Inc. recalls FUSION Dual Pump Packs, REF: OPO73
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and…
FDA · DeviceJohnson & Johnson Surgical Vision, Inc.US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France,…Official notice →US
Agfa Healthcare NV recalls Agfa HealthCare Enterprise Imaging XERO Viewer
There is a software defect that can cause issues with images.
FDA · DeviceAgfa Healthcare NVDistribution throughout US OUS distribution to Australia, Belgium, Brazil,…Official notice →US
Getinge Usa Sales Inc recalls Getinge Flow-e Anesthesia System, Part No. 6887900
Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and…
FDA · DeviceGetinge Usa Sales IncUS nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ,…Official notice →US
Pharmacy Innovations recalls GLUTATHIONE L REDUCED 100MG/ML (2ML NEBVL) INHALATION; GLUTATHIONE L REDUCED 200MG/ML (1ML NEBVL) INHALATION; GLUTATHIONE L REDUCED…
Lack of Assurance of Sterility: FDA inspection revealed insanitary conditions at the facility.
US
Pfizer Inc. recalls Vancomycin Hydrochloride for Injection, USP, 1.5 g/vial, Sterile powder, Single-dose Fliptop Vial, 10 vials per carton, Rx only,…
Presence of Particulate Matter: Glass particulate matter detected in injectable.