Drugs & medicalClass II

Siemens Medical Solutions USA, Inc. recalls SIEMENS Biograph mCT: a) Flow 64-4R Seismic, b) S(40)-3R Seismic, c) S(64)-3R Seismic, d) S(40)-4R Seismic, e) S(64)-4R…

This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.

Read the official notice →
Date
2023-02-22
Country
United States
Brand
Siemens Medical Solutions
Product
SIEMENS Biograph mCT: a) Flow 64-4R Seismic, b) S(40)-3R Seismic, c) S(64)-3R Seismic, d) S(40)-4R Seismic, e) S(64)-4R Seismic, f)…
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Japan, Chile,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Medical Solutions recalls

  1. 2023-05-03
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS one with a hybrid cable- An angiography system developed for diagnostic imaging and interventional procedures. Material…

    Examination room monitor connected with the hybrid cable may have a display issue and no image function is available

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution in the states of FL, SD, PR and the…Official notice →
  3. 2023-04-05
  4. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray system. Material #…

    In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  5. 2023-03-01
  6. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model…

    Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  7. 2023-02-08
  8. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls System Material # Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee multi-purpose 10094139 Artis zee biplane…

    In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution.Official notice →
  9. 2023-02-01
  10. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the…

    In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped

    FDA · DeviceSiemens Medical Solutions USA, IncNationwideOfficial notice →
  11. 2023-01-25
  12. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Luminos Agile Max, Luminos dRF Max, and LUMINOS Lotus Max

    There is potential for the footrest to detach from the patient table during use. Device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system has medical…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distributionOfficial notice →
All 70 Siemens Medical Solutions recalls →

Other drugs & medical recalls in the United States

  1. 2023-03-01
  2. US
  3. US
    Drugs & medicalClass II

    Dr. Reddy's Laboratories, Inc. recalls Tacrolimus Capsules, USP, 0.5 mg, 100-count bottle, Rx Only, Mfd. by Dr. Reddy's Laboratories Limited, Bachupally - 500 090,…

    Presence of Foreign Tablets/Capsules: Presence of one Tacrolimus 1 mg capsule co-mingled in a bottle containing and labeled as Tacrolimus 0.5 mg capsules.

    FDA · DrugDr. Reddy's Laboratories, Inc.Nationwide in the USAOfficial notice →
  4. US
    Drugs & medicalClass III

    Advantice Health, Llc recalls Dermoplast FIRST AID ANTIBACTERIAL SPRAY (Benzethonium chloride 0.2% First aid antiseptic, Benzocaine 20%) Topical analgesic, NET WT. 2.75…

    Subpotent Drug: Low assay observed in one of the two active ingredients during stability testing.

    FDA · DrugAdvantice Health, LlcProduct was distributed via the internet, distributors and retailers nationwide.Official notice →
  5. 2023-02-22
  6. US
    Drugs & medicalClass II

    Conformis, Inc. recalls iTotal Identity Impactor Handle, Model No. 1080-212/ED-07827 REV AE

    Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may loosen and fall into the patient during a procedure.

    FDA · DeviceConformis, Inc.US Nationwide distribution in the states of AL, AZ, CO, FL, GA, KS, KY, LA,…Official notice →
  7. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III Colonovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus,…

    A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination

    FDA · DeviceOlympus Corporation of the AmericasUS Nationwide distribution in the states of CA, FL, GA, NY, OR, VA.Official notice →
  8. US
    Drugs & medicalClass II

    Access Vascular, Inc recalls HydroPICC 4Fr Single Lumen, 80cm Guidewire - Basic Kit, Model Number 80001002

    Select catheter lots were identified to be labeled with incorrect expiration date, where the catheters expire prior to the labeled expiration date of the kit.

    FDA · DeviceAccess Vascular, IncUS Nationwide distribution in the states of CT, IL, IN, MA, NE, TN, and TX.Official notice →