Siemens Medical Solutions USA, Inc recalls ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray system. Material #…
In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.
Read the official notice →- Date
- 2023-04-05
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray system. Material # 10849000.
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2023-05-31
US
Siemens Medical Solutions USA, Inc recalls Luminos Agile Max (VE10, VF10, VF11)
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point,…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - Worldwide DistributionOfficial notice →- 2023-05-10
US
Siemens Medical Solutions USA, Inc recalls ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single…
If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range…
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution.Official notice →- 2023-05-03
US
Siemens Medical Solutions USA, Inc recalls ARTIS one with a hybrid cable- An angiography system developed for diagnostic imaging and interventional procedures. Material…
Examination room monitor connected with the hybrid cable may have a display issue and no image function is available
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution in the states of FL, SD, PR and the…Official notice →- 2023-03-01
US
Siemens Medical Solutions USA, Inc recalls Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model…
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- 2023-02-22
US
Siemens Medical Solutions USA, Inc. recalls SIEMENS Biograph mCT: a) Flow 64-4R Seismic, b) S(40)-3R Seismic, c) S(64)-3R Seismic, d) S(40)-4R Seismic, e) S(64)-4R…
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - US Nationwide and the countries of Japan, Chile,…Official notice →- 2023-02-08
US
Siemens Medical Solutions USA, Inc recalls System Material # Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee multi-purpose 10094139 Artis zee biplane…
In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue…
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution.Official notice →
Other drugs & medical recalls in the United States
- 2023-04-05
US
Stryker Sustainability Solutions recalls Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder…
Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between…
FDA · DeviceStryker Sustainability SolutionsUS: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE,…Official notice →US
Fujifilm Irvine Scientific, Inc. recalls Irvine Scientific Oil for Embryo Culture (Light Mineral Oil) REF Catalog # 9305
Due to the detection of oil toxicity in several lots of Oil for Embryo Culture products.
FDA · DeviceFujifilm Irvine Scientific, Inc.U.S.: AZ, AR, CA, CT, DC, FL, HI, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO,…Official notice →US
Defibtech, LLC recalls DDU-100 Lifeline Automated External Defibrillator (AED)-Indicated for use on victims of sudden cardiac arrest
A component of a sub-assembly used in the affected AEDs had not undergone durability and reliability testing, and if a malfunction occurs, therefore the device could fail to deliver therapy.
US
Iradimed Corporation recalls Iradimed Corporation MRidium MRI 1000 Series Infusion Sets, MRI Infusuui Set - Syringe Adapter Set, REF 1057-50,
Certain Iradimed 1057 Syringe Adapter Sets, used with the MRidium MRI infusion pump, have a problem relating to a defective injection molded part which can limit the effectiveness of the syringe venting function which…
FDA · DeviceIradimed CorporationWorldwide distribution - US Nationwide and the countries of Austria, Canada,…Official notice →US
Elekta Instrument AB recalls Instructions for Use for the following Leksell Stereotactic System and Leksell Neurosurgical Instruments: (1) Part #A2800-26, Backlund…
The incorrect IFU was distributed with the devices.
FDA · DeviceElekta Instrument ABUS Nationwide distribution in the states of AZ, CA, CO, FL, GA, ID, IL, KS,…Official notice →US
Quidel Corporation recalls MicroVue C1-Inhibitor Plus EIA for measuring the amount of functional C1-lnhibitor protein in human plasma or serum, Catalog Number A037,…
Kits with affected microtiter plates have the potential to produce low OD values, causing the run to fail, per the Validation requirements of the Package Insert.