Drugs & medicalClass II

Getinge Usa Sales Inc recalls Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN

The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.

Read the official notice →
Date
2022-11-23
Country
United States
Brand
Getinge Usa Sales
Product
Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN.…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Getinge Usa Sales recalls

  1. 2023-10-11
  2. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Meera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as…

    Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive)…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution.Official notice →
  3. 2023-02-01
  4. US
    Drugs & medicalClass I

    Getinge Usa Sales Inc recalls Getinge Flow-e Anesthesia System, Part No. 6887900

    Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and…

    FDA · DeviceGetinge Usa Sales IncUS nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ,…Official notice →
  5. 2022-12-14
  6. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

    The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power…

    FDA · DeviceGetinge Usa Sales IncUS Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.Official notice →
  7. 2022-08-31
  8. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Servo-air Ventilator System-The Servo-air Ventilator System is: Intended for respiratory support, monitoring and treatment of pediatric…

    Device does not meet regulatory requirements of stability while stationary, which requires mechanical equipment, other than fixed mechanical equipment, that is intended to be used on the floor or on a table shall be…

    FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide and the countries of Australia (AU),…Official notice →
  9. US
    Drugs & medicalClass I

    Getinge Usa Sales Inc recalls Maquet Critical Care AB/Getinge Servo-u Ventilator System, Part Number 6694800 Servo-n Ventilator System, Part Number 6688600 Servo-u MR…

    Ventilators may generate a combination of alarms which may result in loss of communication, technical alarms being triggered, and loss of ventilation. If the device stops ventilating due to this issue, it may lead to…

    FDA · DeviceGetinge Usa Sales IncWorldwide distribution. US nationwide, Poland, Spain, France, Germany, Czech…Official notice →
  10. 2022-07-06
  11. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls VOLISTA StandOP Surgical Light, Product Code/Part Numbers ARD568811901 VLT600SF AIM STP ARD568811911 VLT600SF AIM STP ARD568811961…

    Paint particles may detach and fall in the operating field. In the worst case scenario, infection, organ irritation, hypersensitivity reaction, and tissue granuloma may result if paint particles fall during surgical…

    FDA · DeviceGetinge Usa Sales IncWorldwide distribution - US Nationwide distribution in the states of AR, AZ,…Official notice →
All 17 Getinge Usa Sales recalls →

Other drugs & medical recalls in the United States

  1. 2022-11-23
  2. US
    Drugs & medicalClass II

    Texas Medical Technology Inc. recalls Disposable Surgical Level 3 Gown, Size/Product Code: Large/GWS-03-L-B, XLarge/GWS-03-XL-B, XXLarge/GWS-03-XXL-B

    Surgical gowns recalled due to gown fabric failing to pass the hydrostatic pressure test so the fabric does not provide the protection needed in a level 3 surgical gown, i.e., blood and other fluid may penetrate the…

    FDA · DeviceTexas Medical Technology Inc.US Nationwide distribution in the state of FL.Official notice →
  3. US
    Drugs & medicalClass II

    Steris Corporation recalls GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or…

    Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which…

    FDA · DeviceSteris CorporationUS Nationwide distribution in the states of Arizona, Illinois, Tennessee,…Official notice →
  4. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Simplex HV With Gentamicin CE, 10 PACK-Bone cement with Gentomycin Catalog Number: 6193-1-010

    Simplex HV With Gentamicin CE, 10 PACKS (P/N 6193-1-010) were distributed to customers who had ordered Simplex HV With Gentamicin in US, 10 PACKS (P/N 6195-1-010)

    FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000

    Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients.

    FDA · DeviceAtrium Medical CorporationUS Nationwide distribution in the state of VA.Official notice →
  6. US
    Drugs & medicalClass II

    Sonendo Inc recalls GentleWave System APM Procedure Instrument with Matrix REF GW-APM-PI02

    Procedure Instruments with erroneous unit carton labels.

    FDA · DeviceSonendo IncWorldwide - U.S. Nationwide distribution in the state of OH and the country of…Official notice →
  7. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative…

    Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false…

    FDA · DeviceBecton Dickinson & Co.US Nationwide distribution.Official notice →
See all drugs & medical recalls →