Drugs & medicalClass II

Atrium Medical Corporation recalls Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000

Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients.

Read the official notice →
Date
2022-11-23
Country
United States
Brand
Atrium Medical
Product
Atrium Ocean Wet Suction Water Seal Chest Drain- Product Code: 2002-000
Classification
Class II
Units
US Nationwide distribution in the state of VA.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Atrium Medical recalls

  1. 2023-11-22
  2. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Express Dry Seal Chest Drains (Express): 4000-100N DRAIN, EXPRESS, SINGLE; 4050-100N DRAIN, EXPRESS, BRU

    Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection…

    FDA · DeviceAtrium Medical CorporationWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  3. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac…

    Balloon-to-catheter weld failure that may cause loss of balloon pressure during stent deployment.

    FDA · DeviceAtrium Medical CorporationForeign: Australia, Belgium, Canada, France, Germany, Greece, Malta,…Official notice →
  4. 2023-04-05
  5. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Express Mini 500 Dry Seal Chest Drain, Part Number 16400

    The Instructions for Use (IFU) for the Pneumostat Chest Drain Valve and the Express Mini 500 Dry Seal Chest Drains do not provide sufficient precaution against draining the device for continued use. Additionally, the…

    FDA · DeviceAtrium Medical CorporationWorldwide distribution - US Nationwide and the countries of United Arab…Official notice →
  6. 2022-07-06
  7. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the chest cavity or…

    Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while the case and…

    FDA · DeviceAtrium Medical CorporationAZ IL KS LA MI NY OH OK TX VAOfficial notice →
  8. 2022-05-25
  9. US
    Drugs & medicalClass I

    Atrium Medical Corporation recalls Atrium Advanta V12 Covered Stent System (OUS only). Product Code/Part Number: 85320 85321 85322 85323 85324 85325 85326 85327 85328…

    Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible…

    FDA · DeviceAtrium Medical CorporationForeign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria…Official notice →
  10. 2021-12-01
  11. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls The Ocean Water Seal Chest Drain is a plastic, three bottle vacuum collection device. It consists of a collection chamber that holds…

    Inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient

    FDA · DeviceAtrium Medical CorporationUS Nationwide Distribution Foreign: Bahrain Belgium Brunei Darussalam Canada…Official notice →
All 15 Atrium Medical recalls →

Other drugs & medical recalls in the United States

  1. 2022-11-23
  2. US
    Drugs & medicalClass II

    Steris Corporation recalls GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or…

    Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which…

    FDA · DeviceSteris CorporationUS Nationwide distribution in the states of Arizona, Illinois, Tennessee,…Official notice →
  3. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls Simplex HV With Gentamicin CE, 10 PACK-Bone cement with Gentomycin Catalog Number: 6193-1-010

    Simplex HV With Gentamicin CE, 10 PACKS (P/N 6193-1-010) were distributed to customers who had ordered Simplex HV With Gentamicin in US, 10 PACKS (P/N 6195-1-010)

    FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Getinge Usa Sales Inc recalls Extension Tabletop for Operating Table System 1140, Model Nos. 1140.20F0 and 1140.20AN

    The back rest may unintentionally drop suddenly, resulting in adverse health consequences if a patient were to fall on the floor head first and at a steep angle.

    FDA · DeviceGetinge Usa Sales IncWorldwide - US Nationwide distribution in the states of MA, NJ, SC, and TN.…Official notice →
  5. US
    Drugs & medicalClass II

    Sonendo Inc recalls GentleWave System APM Procedure Instrument with Matrix REF GW-APM-PI02

    Procedure Instruments with erroneous unit carton labels.

    FDA · DeviceSonendo IncWorldwide - U.S. Nationwide distribution in the state of OH and the country of…Official notice →
  6. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibility of most aerobic and facultative…

    Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false…

    FDA · DeviceBecton Dickinson & Co.US Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Hill-Rom Progressa Bed System

    The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur…

    FDA · DeviceBaxter Healthcare CorporationWorldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, HI,…Official notice →
See all drugs & medical recalls →