Drugs & medicalClass II

Covidien, LLC recalls Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.

Read the official notice →
Date
2023-07-05
Country
United States
Brand
Covidien
Product
Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS
Classification
Class II
Units
International distribution in the country of Taiwan.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Covidien recalls

  1. 2023-10-04
  2. US
    Drugs & medicalClass II

    Covidien LP recalls Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm

    Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.

    FDA · DeviceCovidien LPWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  3. 2023-08-23
  4. US
    Drugs & medicalClass I

    Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP

    Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully,…

    FDA · DeviceCovidien LPUS Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain…Official notice →
  5. 2023-08-09
  6. US
    Drugs & medicalClass II

    Covidien, LLC recalls Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM

    Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction,…

    FDA · DeviceCovidien, LLCWorldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the…Official notice →
  7. 2023-05-03
  8. US
    Drugs & medicalClass II

    Covidien LP recalls VersaOne Reusable Positioning Cannula 11mm Standard, Model Number RC11STS

    There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.

    FDA · DeviceCovidien LPWorldwide - US Nationwide distribution in the states of NC and the countries…Official notice →
  9. 2023-04-26
  10. US
    Drugs & medicalClass I

    Covidien recalls Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF,…

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories.…

    FDA · DeviceCovidienWorldwide - US Nationwide distribution including in the states of PR, CA, FL,…Official notice →
  11. 2023-04-19
  12. US
    Drugs & medicalClass II

    Covidien, LP recalls Step Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer, Product Number S101010

    Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of…

    FDA · DeviceCovidien, LPWorldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA,…Official notice →
All 53 Covidien recalls →

Other drugs & medical recalls in the United States

  1. 2023-07-05
  2. US
    Drugs & medicalClass III

    Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…

    Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…

    FDA · DeviceRandox Laboratories Ltd.CA, IL, VA, WA, WI including PROfficial notice →
  3. US
    Drugs & medicalClass II

    Tornier, Inc recalls AEQUALIS FLEX REVIVE Assembly Screw 0mm Short, Model Number ARS655118

    A comingle/swap resulted in the labeling and laser marking on the product to state it is a Aequalis Flex Revive Assembly Screw 0mm Standard (ARS655101) but it was a Aequalis Flex Revive Assembly Screw 0mm Short…

    FDA · DeviceTornier, IncWorldwide distribution - US Nationwide and the countries of Australia, Canada,…Official notice →
  4. US
    Drugs & medicalClass II

    Limacorporate S.p.A recalls REF 1975.14.800, Prima TT Genoid Modular Reverse TT Baseplate, Sterile R

    Due to manufacturing issue, their is a potential for peripheral holes to be out of specification.

    FDA · DeviceLimacorporate S.p.AUS Nationwide distribution in the states of MI, IL, ND, MO, OK.Official notice →
  5. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls NaviCare Patient Safety V3.9.200 to 3.9.600, used with the Centrella Bed

    Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned…

    FDA · DeviceBaxter Healthcare CorporationUS Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.Official notice →
  6. US
    Drugs & medicalClass II

    Permobil recalls PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D,…

    When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be…

    FDA · DevicePermobilWorldwide distribution.Official notice →
  7. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls Design Options¿ / Epidural anesthesia kit (10 count carton)

    Kits were assembled with the incorrect Filter Straw.

    FDA · DeviceB. Braun Medical, Inc.US Nationwide distribution.Official notice →
See all drugs & medical recalls →