Drugs & medicalClass II

Covidien, LLC recalls Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM

Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction,…

Read the official notice →
Date
2023-08-09
Country
United States
Brand
Covidien
Product
Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Covidien recalls

  1. 2023-10-11
  2. US
    Drugs & medicalClass II

    Covidien recalls Puritan Bennett Cuff Pressure Manager, REF: 180-03

    Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from…

    FDA · DeviceCovidienUS: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El…Official notice →
  3. 2023-10-04
  4. US
    Drugs & medicalClass II

    Covidien LP recalls Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm

    Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.

    FDA · DeviceCovidien LPWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  5. 2023-08-23
  6. US
    Drugs & medicalClass I

    Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP

    Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully,…

    FDA · DeviceCovidien LPUS Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain…Official notice →
  7. 2023-07-05
  8. US
    Drugs & medicalClass II

    Covidien, LLC recalls Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

    Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.

    FDA · DeviceCovidien, LLCInternational distribution in the country of Taiwan.Official notice →
  9. 2023-05-03
  10. US
    Drugs & medicalClass II

    Covidien LP recalls VersaOne Reusable Positioning Cannula 11mm Standard, Model Number RC11STS

    There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.

    FDA · DeviceCovidien LPWorldwide - US Nationwide distribution in the states of NC and the countries…Official notice →
  11. 2023-04-26
  12. US
    Drugs & medicalClass I

    Covidien recalls Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF,…

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories.…

    FDA · DeviceCovidienWorldwide - US Nationwide distribution including in the states of PR, CA, FL,…Official notice →
All 53 Covidien recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-09
  2. US
    Drugs & medicalClass II

    Materialise N.V. recalls ProPlan CMF Patient-Specific Guide, Mandible guides (3A & 3B)-Intended to guide the marking of bone and/or guide surgical instruments in…

    Wrong versions of guides 3A and 3B (Patient Specific) were shipped prior to the re-design request

    FDA · DeviceMaterialise N.V.US Nationwide distribution in the state of KY.Official notice →
  3. US
    Drugs & medicalClass II

    Scican Ltd. recalls STATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by…

    The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the…

    FDA · DeviceScican Ltd.US Nationwide Distribution and in the country of CanadaOfficial notice →
  4. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20. X-ray Image guided Therapy System.

    The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  5. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Impella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323

    A higher than expected rate of thrombus formation or deposition has been observed. The risk is for thrombus formations or deposits on indwelling central venous lines or cannulas may break free and enter into the…

    FDA · DeviceAbiomed, Inc.US Nationwide distribution including in the states of AR, CA, FL, GA, LA, MA,…Official notice →
  6. US
    Drugs & medicalClass II

    Hamamatsu Corporation recalls NanoZoomer S360MD Slide scanner system-automated system for creating, viewing, and managing digital slides. The NanoZoomer System creates…

    If the wrong barcode information is recorded in the NDPi file, a pathologist may refer to the whole slide image (WSI) of a different patient.

    FDA · DeviceHamamatsu CorporationNYOfficial notice →
  7. US
    Drugs & medicalClass II

    GE OEC Medical Systems, Inc recalls OEC Flexiview 8800 Systems with 9-inch Image Intensifier

    Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which…

    FDA · DeviceGE OEC Medical Systems, IncWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →
See all drugs & medical recalls →