Abiomed, Inc. recalls Impella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323
A higher than expected rate of thrombus formation or deposition has been observed. The risk is for thrombus formations or deposits on indwelling central venous lines or cannulas may break free and enter into the…
Read the official notice →- Date
- 2023-08-09
- Country
- United States
- Brand
- Abiomed
- Product
- Impella RP Flex with SmartAssist intracardiac microaxial blood pump, Product Number 1000323
- Classification
- Class I
- Units
- US Nationwide distribution including in the states of AR, CA, FL, GA, LA, MA,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Abiomed recalls
- 2024-01-17
US
Abiomed, Inc. recalls Abiomed Kit, 14Fr Introducer, 13cm&25cm, Sterile Part of Pump Set (0048-0040, 0048-0045, 1000080)-Introducer(s) are intended for…
Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of…
FDA · DeviceAbiomed, Inc.Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX…Official notice →- 2024-01-10
US
Abiomed, Inc. recalls Impella Connect, software portal for online device management that allows clinicians and Impella support staff to remotely view Impella…
The following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature;…
- 2023-12-13
US
Abiomed, Inc. recalls Automated Impella Controller (AIC), product number 0042-040-US with software versions V8.4 and V8.4.1.
Pump not detected as connected to controller due to software issue.
- 2023-07-26
US
Abiomed, Inc. recalls Impella LD intravascular micro axial blood pump, Product Number 005082
There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has…
- 2023-07-19
US
Abiomed, Inc. recalls Abiomed 14Fr Low Profile Introducer Set, Product Number 1000457
There was one complaint that the sidearm of the device detached during explant of a pump. When the sidearms are subjected to heat, humidity and articulation, a silicone oil contaminant that was unintentionally…
- 2023-05-24
US
Abiomed, Inc. recalls Impella 5.5 with SmartAssist catheter is an intravascular micro axial blood pump that supports a patient s circulatory system, Product…
The Impella 5.5 pump experienced a heightened complaint rate for purge leaks.
FDA · DeviceAbiomed, Inc.US Nationwide. Japan, Saudi Arabia, Kuwait, Italy, Spain, Switzerland,…Official notice →
Other drugs & medical recalls in the United States
- 2023-08-09
US
Scican Ltd. recalls STATCLAVE G4 Chamber Autoclave -Tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by…
The STATCLAVE G4 is a dynamic air removal (prevacuum) tabletop steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the…
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20. X-ray Image guided Therapy System.
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Covidien, LLC recalls Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction,…
FDA · DeviceCovidien, LLCWorldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the…Official notice →US
Hamamatsu Corporation recalls NanoZoomer S360MD Slide scanner system-automated system for creating, viewing, and managing digital slides. The NanoZoomer System creates…
If the wrong barcode information is recorded in the NDPi file, a pathologist may refer to the whole slide image (WSI) of a different patient.
US
GE OEC Medical Systems, Inc recalls OEC Flexiview 8800 Systems with 9-inch Image Intensifier
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which…
FDA · DeviceGE OEC Medical Systems, IncWorldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →US
Maquet Cardiovascular, LLC recalls ACROBAT SUV Vacuum Off-Pump System, Part Number C-OM-9000S. Component of a chest stabilization system for cardiac and chest surgery.
The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device…
FDA · DeviceMaquet Cardiovascular, LLCWorldwide distribution - US Nationwide and the countries of Bangladesh,…Official notice →