Covidien LP recalls Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.
Read the official notice →- Date
- 2023-10-04
- Country
- United States
- Brand
- Covidien
- Product
- Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Covidien recalls
- 2023-11-08
US
Covidien recalls McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-017-000, Sterile EO, Rx Only
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
FDA · DeviceCovidienUS: AK AL AR AZ CA CO CT DE FL GA IA IL IN KS LA MA MD MI MO MT NC ND NH NJ NM…Official notice →US
Covidien recalls Stolen McGRATH MAC Video Laryngoscopes, REF: 301-000-000
Defective video laryngoscopes that did not pass quality tests established for the product, that were designated for destruction, were stolen, and subsequently distributed for sale. Defective device use may result in…
- 2023-10-11
US
Covidien recalls Puritan Bennett Cuff Pressure Manager, REF: 180-03
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from…
FDA · DeviceCovidienUS: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El…Official notice →- 2023-08-23
US
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully,…
FDA · DeviceCovidien LPUS Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain…Official notice →- 2023-08-09
US
Covidien, LLC recalls Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction,…
FDA · DeviceCovidien, LLCWorldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the…Official notice →- 2023-07-05
US
Covidien, LLC recalls Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS
Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.
Other drugs & medical recalls in the United States
- 2023-10-04
US
Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital…
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber…
FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of NJ, WI.Official notice →US
Inmar Supply Chain Solutions, LLC recalls Various Medical Products stored on pallets, including but not limited to bandages, tampons, condoms, blood glucose monitors,…
Various medical devices potentially exposed to rodent and rodent activity. In addition, medical devices may have been subjected to temperatures in excess of storage requirements on product labeling.
US
Exocad GmbH recalls exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the…
A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a guided surgery treatment approach is…
FDA · DeviceExocad GmbHWorldwide distribution - US Nationwide and the country of Canada.Official notice →US
Philips Respironics, Inc. recalls Philips Respironics V60/V60 Plus Ventilator, PCBA Part Number: 453561544461 V60 Device Part Numbers 1053617 (US) and 1137276…
Power Management PCBAs may malfunction, causing a power failure leading to ventilator loss of function.
FDA · DevicePhilips Respironics, Inc.US Distribution to states of: OH, CA, SC, OR; and OUS country of: Philippines.Official notice →US
WOM World Of Medicine AG recalls Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to…
The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached…
FDA · DeviceWOM World Of Medicine AGNationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany,…Official notice →US
Ambu Inc. recalls Ambu aView 2 A Monitor-The Ambu aView 2 Advance is intended to display live imaging data from compatible Ambu visualization devices,…
Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames