Drugs & medicalClass II

Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital…

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber…

Read the official notice →
Date
2023-10-04
Country
United States
Brand
Fresenius Kabi
Product
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital…
Classification
Class II
Units
US Nationwide distribution in the states of NJ, WI.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Fresenius Kabi recalls

  1. 2024-03-20
  2. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS); LVP-SW-0004

    Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to…

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution to the states of CA, CO, NJ, WI, UT.Official notice →
  3. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004

    Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable…

    FDA · DeviceFresenius Kabi USA, LLCUS Nationwide distribution in the states of WI and NJ.Official notice →
  4. 2023-12-27
  5. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump LVP-0004

    The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of…

    FDA · DeviceFresenius Kabi USA, LLCUS distribution to states of: CA, CO, NJ, UT, and WI.Official notice →
  6. 2023-04-19
  7. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump, Model No. LVP-0004

    Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.

    FDA · DeviceFresenius Kabi USA, LLCDomestic: CA, CO, NJ, WI, & UT. No foreign distribution.Official notice →
  8. 2023-01-25
  9. US
    Drugs & medicalClass II

    Fresenius Kabi USA, LLC recalls Sensorcaine (Bupivacaine HCl and Epinephrine Injection, USP) with Epinephrine 1:200,000 (as bitartrate), 0.5%, 250 mg per 50 mL (5 mg per…

    Subpotent Drug: Testing results below the defined limit for the epinephrine portion of this product.

    FDA · DrugFresenius Kabi USA, LLCNationwide in the USAOfficial notice →
  10. 2022-10-12
  11. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0

    The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does…

    FDA · DeviceFresenius Kabi USA, LLCDomestic distribution to NJ and WI.Official notice →
All 31 Fresenius Kabi recalls →

Other drugs & medical recalls in the United States

  1. 2023-10-04
  2. US
    Drugs & medicalClass II

    Waldemar Link GmbH & Co. KG (Mfg Site) recalls MobileLink Acetabular Cup System, Shell/Insert Adapter for Shell 74-80mm, Insert Size F, +12mm Offset 20 degrees…

    Product is labeled with incorrect offset.

    FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Domestic distribution to AL, NJ, NV.Official notice →
  3. US
    Drugs & medicalClass II

    3M Company - Health Care Business recalls 3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam…

    Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain…

    FDA · Device3M Company - Health Care BusinessWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  4. US
    Drugs & medicalClass II

    J&T Health and Wellness LLC recalls ProSun International Luxura Vegaz, Model 52X; Ultraviolet Tanning Device

    In some cases the communication isolator installed in the Model 52X can cause a disturbance in the glass-filter safety circuit.

    FDA · DeviceJ&T Health and Wellness LLCUSOfficial notice →
  5. US
    Drugs & medicalClass II

    Inmar Supply Chain Solutions, LLC recalls Various Medical Products stored on pallets, including but not limited to bandages, tampons, condoms, blood glucose monitors,…

    Various medical devices potentially exposed to rodent and rodent activity. In addition, medical devices may have been subjected to temperatures in excess of storage requirements on product labeling.

    FDA · DeviceInmar Supply Chain Solutions, LLCUS Nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass II

    Exocad GmbH recalls exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the…

    A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a guided surgery treatment approach is…

    FDA · DeviceExocad GmbHWorldwide distribution - US Nationwide and the country of Canada.Official notice →
  7. US
    Drugs & medicalClass II

    Covidien LP recalls Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm

    Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.

    FDA · DeviceCovidien LPWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
See all drugs & medical recalls →