Hologic, Inc recalls Sertera 14 Gauge Biopsy Device-intended to obtain percutaneous core biopsy samples from soft tissue and tumors of the breast REF: SERTERA-14
The inner needle of the Sertera device is either detaching or discharging and may cause delay in surgery
Read the official notice →- Date
- 2023-07-05
- Country
- United States
- Brand
- Hologic
- Product
- Sertera 14 Gauge Biopsy Device-intended to obtain percutaneous core biopsy samples from soft tissue and tumors of the breast REF: SERTERA-14
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Hologic recalls
- 2024-05-15
US
Hologic, Inc recalls BioZorb Marker-BioZorb 3D Bioabsorbable Marker-Intended Use radiographic marking of sites in soft tissue. 1) F0405 BioZorb Marker 4cm x…
Safety Information: Complaints describe complications/adverse events that include pain, infection, rash, device migration, device erosion, seroma, discomfort, or other complications from feeling the device in the…
- 2024-04-03
US
Hologic, Inc. recalls Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter…
Use of assay along with plasma specimen may result in error code which would invalidate sample.
FDA · DeviceHologic, Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL,…Official notice →- 2024-01-31
US
Hologic, Inc. recalls PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system which fully automates all steps…
Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive.
FDA · DeviceHologic, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AR,…Official notice →- 2023-03-01
US
Hologic, Inc. recalls Hologic 3Dimensions/Selenia Dimensions Models: RM-SDM-00001-3D SDM-00001-2D SDM-00001-3D SDM-05000-3D3 SDM-05000-2DC SDM-00001-M3D…
C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual
- 2022-10-12
US
Hologic, Inc recalls The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is…
Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is…
FDA · DeviceHologic, IncUS Nationwide distribution including in the states of FL, IL, MD, NJ, AZ, TX,…Official notice →- 2022-04-13
US
Hologic, Inc recalls Panther Fusion GBS Assay Cartridges 96 Tests, Catalog Number: PRD-04484. in vitro diagnostic test for group b streptococcus.
Samples, run on a GBS Assay, that have analyte concentrations that are at or near the limit of detection may produce false negative results.
FDA · DeviceHologic, IncUS: TX, TN, NY, NV, IN, AR, CA, CT, IL, NH, PA, WA, FL, OR, WV, VA, UT, MA,…Official notice →
Other drugs & medical recalls in the United States
- 2023-07-05
US
Remel, Inc recalls Remel Haemophilus Test Medium (Agar) (150mm), 10/pk, REF R04033; and Remel Haemophilus Test Medium (Agar)(100mm), 10/pk, REF 01503. For In…
The test medium may not perform as intended.
FDA · DeviceRemel, IncDistribution was made nationwide, including Puerto Rico. There was no foreign…Official notice →US
Implant Direct Sybron Manufacturing LLC recalls LEGACY SMARTBASE, NE 3.0D 1MML, REF: 8730-81NE; LEGACY SMARTBASE, NE 3.0D 2MML, REF 8730-82NE; LEGACY SMARTBASE NARROW…
Non-engaging abutments were distributed without FDA clearance; the performance characteristics have not yet been adequately established.
FDA · DeviceImplant Direct Sybron Manufacturing LLCUS Nationwide distribution in the states of FL, ID, TX, CA.Official notice →US
Medtronic Neuromodulation recalls Vanta with AdaptiveStim Implantable Neurostimulator, Model 977006; spinal chord stimulation
A cardioversion procedure may damage the electronics in the Vanta INS device, making the INS unresponsive and non-functional. As of April 19, 2023, Medtronic has received a total of two complaints concerning this issue…
US
DRE Medical Group Inc recalls Philips Respironics V60 Ventilator-Intended to mechanically control or assist patient breathing by delivering a predetermined percentage…
Devices were imported with Japanese Language and NIST Oxygen Gas Fittings and retrofitted, may pose a serious health risk as the gas pathway of these devices may be contaminated.
US
Luminex Corporation recalls VERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-022
It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result…
US
Randox Laboratories Ltd. recalls Immunoassay Premium Level 3-In vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of…
Delay in reporting results due to Quality Controls running high outside of range. Target value and range for ACTH in Immunoassay Premium Level 3, IA2640, lot 2031EC and Immunoassay Premium Control Tri Level kit batch…