Drugs & medicalClass I

Hologic, Inc recalls BioZorb Marker-BioZorb 3D Bioabsorbable Marker-Intended Use radiographic marking of sites in soft tissue. 1) F0405 BioZorb Marker 4cm x…

Safety Information: Complaints describe complications/adverse events that include pain, infection, rash, device migration, device erosion, seroma, discomfort, or other complications from feeling the device in the…

Read the official notice →
Date
2024-05-15
Country
United States
Brand
Hologic
Product
BioZorb Marker-BioZorb 3D Bioabsorbable Marker-Intended Use radiographic marking of sites in soft tissue. 1) F0405 BioZorb Marker 4cm x…
Classification
Class I
Units
US Nationwide Distribution
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Hologic recalls

  1. 2024-12-04
  2. US
    Drugs & medicalClass I

    Hologic, Inc recalls BioZorb Marker. Radiographic soft tissue marker.

    Firm received reports of adverse events/complications that include serious injuries, such as pain, infection, rash, device migration, device erosion, seroma, and discomfort.

    FDA · DeviceHologic, IncDomestic: US Nationwide Distribution.Official notice →
  3. 2024-10-02
  4. US
    Drugs & medicalClass II

    Hologic Inc recalls Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as…

    Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may…

    FDA · DeviceHologic IncWorldwide distribution - US Nationwide and the countries of Belgium, Canada,…Official notice →
  5. 2024-07-24
  6. US
    Drugs & medicalClass II

    Hologic Inc recalls Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A, Horizon-W, Horizon-WI, Horizon-C, Horizon-CI, Horizon-A-CN, Horizon-W-CN,…

    Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the…

    FDA · DeviceHologic IncWorldwide distribution - US Nationwide and the countries of India, Australia,…Official notice →
  7. 2024-04-03
  8. US
    Drugs & medicalClass II

    Hologic, Inc. recalls Aptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent, Enzyme Reagent, Promoter…

    Use of assay along with plasma specimen may result in error code which would invalidate sample.

    FDA · DeviceHologic, Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, ID, IL,…Official notice →
  9. 2024-01-31
  10. US
    Drugs & medicalClass II

    Hologic, Inc. recalls PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system which fully automates all steps…

    Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive.

    FDA · DeviceHologic, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK, AR,…Official notice →
  11. 2023-07-05
  12. US
    Drugs & medicalClass II

    Hologic, Inc recalls Sertera 14 Gauge Biopsy Device-intended to obtain percutaneous core biopsy samples from soft tissue and tumors of the breast REF: SERTERA-14

    The inner needle of the Sertera device is either detaching or discharging and may cause delay in surgery

    FDA · DeviceHologic, IncUS Nationwide distribution.Official notice →
All 21 Hologic recalls →

Other drugs & medical recalls in the United States

  1. 2024-05-15
  2. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used…

    Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.

    FDA · DeviceRandox Laboratories Ltd.Domestic: Nationwide Distribution.Official notice →
  3. US
    Drugs & medicalClass II

    SPINEART SA recalls PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55…

    Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but…

    FDA · DeviceSPINEART SAU.S. Nationwide distribution in the states of CA, and KY. Not provided due to…Official notice →
  4. US
    Drugs & medicalClass II

    Philips Ultrasound, Inc. recalls 5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300…

    Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer…

    FDA · DevicePhilips Ultrasound, Inc.Worldwide - US Nationwide distribution including in the states of CA, CO, CT,…Official notice →
  5. US
    Drugs & medicalClass I

    Vyaire Medical, Inc. recalls Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used in the Vyntus CPX.…

    Potential of the nozzle separating during patient use.

    FDA · DeviceVyaire Medical, Inc.US nationwide distribution.Official notice →
  6. US
    Drugs & medicalClass I

    Tandem Diabetes Care, Inc. recalls t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology

    During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result…

    FDA · DeviceTandem Diabetes Care, Inc.US nationwide distribution including Puerto Rico.Official notice →
  7. US
    Drugs & medicalClass II

    SUN PHARMACEUTICAL INDUSTRIES INC recalls Amphotericin B Liposome for Injection, 50mg vials, Rx only, Distributed by: Sun Pharmaceutical Industries, Inc. Cranbury, NJ…

    Out of specification for assay

    FDA · DrugSUN PHARMACEUTICAL INDUSTRIES INCNationwide within the United StatesOfficial notice →
See all drugs & medical recalls →