Drugs & medicalClass II

Medtronic Sofamor Danek USA, Inc recalls CD HORIZON Spinal System Solera Awl Tip Tap, REF NAV2006K

Incorrect Laser Marking - Solera Awl Tip Tap

Read the official notice →
Date
2023-08-16
Country
United States
Brand
Medtronic Sofamor Danek
Product
CD HORIZON Spinal System Solera Awl Tip Tap, REF NAV2006K
Classification
Class II
Units
US Nationwide distribution in the states of FL, CO, IN, SC, TX, UT.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Medtronic Sofamor Danek recalls

  1. 2025-12-03
  2. US
    Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers: 1. SPACER 6068073 CATALYFT…

    Expandable interbody system cage may loose height or collapse (loss of lordosis), which could result in migration, subsidence and/or neurological injury. Therefore, surgical technique and instructions for use will be…

    FDA · DeviceMedtronic Sofamor Danek USA IncWorldwide distribution - US Nationwide and Puerto Rico. The countries of…Official notice →
  3. 2023-11-01
  4. US
    Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls Grafton DBM (Flex), Model Numbers: T42100, Demineralized Bone Matrix

    Specific serial numbers of GraftonTM DBM product due to the potential that a step during the inspection of the outer Tyvek pouch may not have been performed according to requirements. Based on this, batches may have…

    FDA · DeviceMedtronic Sofamor Danek USA IncUS, Colombia, S. Korea, New Zealand, India, TaiwanOfficial notice →
  5. 2023-09-20
  6. US
    Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls Magnifuse Bone Graft, Large, Size: 14mm x 25mm, REF 7509014, Human Tissue, Demineralized Bone Matrix, sterile.

    The incorrect product labeling was applied to the product indicating the wrong size of product.

    FDA · DeviceMedtronic Sofamor Danek USA IncUS Nationwide distribution in the states of ID, IN, MN, NC, NY, OH, OR, WA,…Official notice →
  7. 2023-07-19
  8. US
    Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls Medtronic Catalyft LS Expandable Interbody System Implant Kit, Anterior Standalone Cage, REF SPS03188, spinal implant kit

    There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral…

    FDA · DeviceMedtronic Sofamor Danek USA IncUS Nationwide distribution in the states of IN, WI, GA, OR, AR, OH, TN, TX,…Official notice →
  9. 2023-05-17
  10. US
    Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal…

    Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.

    FDA · DeviceMedtronic Sofamor Danek USA IncUS: IN, CT, NEOfficial notice →
  11. 2023-04-26
  12. US
    Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and…

    Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

    FDA · DeviceMedtronic Sofamor Danek USA IncWorldwide distribution - US Nationwide and the country of Canada.Official notice →
All 9 Medtronic Sofamor Danek recalls →

Other drugs & medical recalls in the United States

  1. 2023-08-16
  2. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls Total Hip Kit, REF DYNJ903275S; surgical convenience kit

    The prep solution included in the Total Hip Kit expires prior to the expiration date of the Kit.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the state of Illinois.Official notice →
  3. US
    Drugs & medicalClass II

    Medtronic Neuromodulation recalls Medtronic InterStim Ground Pad REF 041826

    Lot of InterStim Self-Adhesive Ground Pad is being recalled due to the product being labeled with the incorrect expiration date.

    FDA · DeviceMedtronic NeuromodulationInternational distribution to the country of Iran.Official notice →
  4. US
    Drugs & medicalClass II

    Scholly Fiberoptic Gmbh recalls UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into…

    There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this…

    FDA · DeviceScholly Fiberoptic GmbhNationwide Foreign: Japan, South KoreaOfficial notice →
  5. US
    Drugs & medicalClass II

    Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)

    Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after…

    FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material…

    Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable…

    FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  7. US
    Drugs & medicalClass II

    ORTHO-CLINICAL DIAGNOSTICS recalls VITROS Immunodiagnostic Products Troponin I ES Reagent Pack

    Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.

    FDA · DeviceORTHO-CLINICAL DIAGNOSTICSWorldwide distribution - US Nationwide and the country of Canada.Official notice →
See all drugs & medical recalls →