Drugs & medicalClass II

Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)

Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…

Read the official notice →
Date
2023-11-22
Country
United States
Brand
Siemens Medical Solutions
Product
Mobilett Mira wireless (VA20) mobile x-ray system(s)
Classification
Class II
Units
US Nationwide - Worldwide
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Siemens Medical Solutions recalls

  1. 2024-04-24
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137…

    Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break .…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  3. 2024-01-17
  4. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Sensis Vibe Hemo

    The possibility of the Sensis documentation functionality application to crash.

    FDA · DeviceSiemens Medical Solutions USA, IncDomestic: CA, CT, & FL.Official notice →
  5. 2023-12-20
  6. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100

    Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL…Official notice →
  7. 2023-11-08
  8. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls ***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and…

    Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure

    FDA · DeviceSiemens Medical Solutions USA, IncNationwideOfficial notice →
  9. 2023-10-11
  10. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica…

    In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  11. 2023-09-20
  12. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 2.0 ultrasound systems, REF: 11503314

    If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6…

    FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide - US Nationwide distribution including in the states of IL, WI, CA,…Official notice →
All 70 Siemens Medical Solutions recalls →

Other drugs & medical recalls in the United States

  1. 2023-11-22
  2. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient…

    The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.

    FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the states of Arkansas, Florida,…Official notice →
  3. US
    Drugs & medicalClass II

    Philips North America Llc recalls Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and…

    For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect…

    FDA · DevicePhilips North America LlcNationwide Foreign: Austria Belgium Canada China Estonia France Germany…Official notice →
  4. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac…

    Balloon-to-catheter weld failure that may cause loss of balloon pressure during stent deployment.

    FDA · DeviceAtrium Medical CorporationForeign: Australia, Belgium, Canada, France, Germany, Greece, Malta,…Official notice →
  5. US
  6. US
    Drugs & medicalClass II

    Omega & Delta Co., Inc. recalls Equate Hand Sanitizer (Ethyl Alcohol 62%), 7.5 FL.OZ. (221 mL) UPC 6 81131 05961 9

    Sub Potent and Super Potent Product: During an impact assessment of the DOP Hand Sanitizer product, it was identified that several lots distributed did not comply with the finished good acceptance criteria established.

    FDA · DrugOmega & Delta Co., Inc.Distributed in Puerto Rico only.Official notice →
  7. 2023-11-15
  8. US
    Drugs & medicalClass II

    Maquet Medical Systems USA recalls Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)

    It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual…

    FDA · DeviceMaquet Medical Systems USAUS Nationwide.Official notice →