Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)
Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…
Read the official notice →- Date
- 2023-11-22
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Mobilett Mira wireless (VA20) mobile x-ray system(s)
- Classification
- Class II
- Units
- US Nationwide - Worldwide
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2024-04-24
US
Siemens Medical Solutions USA, Inc recalls Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137…
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break .…
- 2024-01-17
US
Siemens Medical Solutions USA, Inc recalls Sensis Vibe Hemo
The possibility of the Sensis documentation functionality application to crash.
- 2023-12-20
US
Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL…Official notice →- 2023-11-08
US
Siemens Medical Solutions USA, Inc recalls ***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and…
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
- 2023-10-11
US
Siemens Medical Solutions USA, Inc recalls (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica…
In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential…
- 2023-09-20
US
Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 2.0 ultrasound systems, REF: 11503314
If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6…
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide - US Nationwide distribution including in the states of IL, WI, CA,…Official notice →
Other drugs & medical recalls in the United States
- 2023-11-22
US
Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient…
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the states of Arkansas, Florida,…Official notice →US
Philips North America Llc recalls Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and…
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect…
FDA · DevicePhilips North America LlcNationwide Foreign: Austria Belgium Canada China Estonia France Germany…Official notice →US
Atrium Medical Corporation recalls Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac…
Balloon-to-catheter weld failure that may cause loss of balloon pressure during stent deployment.
FDA · DeviceAtrium Medical CorporationForeign: Australia, Belgium, Canada, France, Germany, Greece, Malta,…Official notice →US
US
Omega & Delta Co., Inc. recalls Equate Hand Sanitizer (Ethyl Alcohol 62%), 7.5 FL.OZ. (221 mL) UPC 6 81131 05961 9
Sub Potent and Super Potent Product: During an impact assessment of the DOP Hand Sanitizer product, it was identified that several lots distributed did not comply with the finished good acceptance criteria established.
- 2023-11-15
US
Maquet Medical Systems USA recalls Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual…