Drugs & medicalClass II

Maquet Medical Systems USA recalls Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)

It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual…

Read the official notice →
Date
2023-11-15
Country
United States
Brand
Maquet Medical Systems
Product
Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)
Classification
Class II
Units
US Nationwide.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Maquet Medical Systems recalls

  1. 2023-12-27
  2. US
    Drugs & medicalClass III

    Maquet Medical Systems USA recalls CARDIOHELP-i Heart Lung Support System: Blood oxygenation and carbon dioxide removal system used to pump blood through the…

    CARDIOHELP-i System was not properly tested to measure leakage current

    FDA · DeviceMaquet Medical Systems USANationwide-WW Distribution to Argentina, Australia, Austria, Azerbaijan,…Official notice →
  3. 2023-12-20
  4. US
    Drugs & medicalClass II

    Maquet Medical Systems USA recalls CARDIOHELP-i Heart Lung Support System- A blood oxygenation and carbon dioxide removal system used to pump blood through the…

    Certain factory settings for the device were incorrectly described in the IFU-related to the warning limits for PVen and PAux, and the default settings for the venous bubble sensor (VBS) and the automatic lock in the…

    FDA · DeviceMaquet Medical Systems USAWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  5. 2023-12-13
  6. US
    Drugs & medicalClass I

    Maquet Medical Systems USA recalls CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide…

    Product removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and…

    FDA · DeviceMaquet Medical Systems USAUS Nationwide - Worldwide Distribution: Domestic distribution to: AL, CA, DC,…Official notice →
  7. 2023-06-07
  8. US
    Drugs & medicalClass II

    Maquet Medical Systems USA recalls QUADROX -i Neonatal, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 10000-USA 701070412 HMO 11000-USA 701070416

    Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis,…

    FDA · DeviceMaquet Medical Systems USANationwide domestic distributionOfficial notice →
  9. US
    Drugs & medicalClass II

    Maquet Medical Systems USA recalls QUADROX-i Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. HMO 30000-USA 701070384 HMO 31000-USA 701070388

    1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may…

    FDA · DeviceMaquet Medical Systems USADomestic distribution nationwide.Official notice →
  10. 2023-03-08
  11. US
    Drugs & medicalClass II

    Maquet Medical Systems USA recalls BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0, Product Code 70106.9078.…

    Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.

    FDA · DeviceMaquet Medical Systems USAWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
All 14 Maquet Medical Systems recalls →

Other drugs & medical recalls in the United States

  1. 2023-11-22
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)

    Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - WorldwideOfficial notice →
  3. US
  4. US
    Drugs & medicalClass II

    Omega & Delta Co., Inc. recalls Equate Hand Sanitizer (Ethyl Alcohol 62%), 7.5 FL.OZ. (221 mL) UPC 6 81131 05961 9

    Sub Potent and Super Potent Product: During an impact assessment of the DOP Hand Sanitizer product, it was identified that several lots distributed did not comply with the finished good acceptance criteria established.

    FDA · DrugOmega & Delta Co., Inc.Distributed in Puerto Rico only.Official notice →
  5. 2023-11-15
  6. US
    Drugs & medicalClass II

    Carestream Health, Inc. recalls The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. the system…

    There is a potential for unexpected failure of the electrical components within the Carestream Health Inc. generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self…

    FDA · DeviceCarestream Health, Inc.US Nationwide. Global Distribution.Official notice →
  7. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls EVIS EXERA III Gastrointestinal Videoscope

    Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.

    FDA · DeviceOlympus Corporation of the AmericasNationwide Distribution.Official notice →
  8. US