Siemens Medical Solutions USA, Inc recalls (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica…
In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential…
Read the official notice →- Date
- 2023-10-11
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica MAGNETOM Spectra…
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2023-12-20
US
Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL…Official notice →- 2023-11-22
US
Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)
Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…
- 2023-11-08
US
Siemens Medical Solutions USA, Inc recalls ***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and…
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
- 2023-09-20
US
Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 2.0 ultrasound systems, REF: 11503314
If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6…
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide - US Nationwide distribution including in the states of IL, WI, CA,…Official notice →- 2023-08-16
US
Siemens Medical Solutions USA, Inc recalls ARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material…
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable…
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →- 2023-08-09
US
Siemens Medical Solutions USA, Inc. recalls The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor…
During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the…
FDA · DeviceSiemens Medical Solutions USA, Inc.Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no…Official notice →
Other drugs & medical recalls in the United States
- 2023-10-11
US
Covidien recalls Puritan Bennett Cuff Pressure Manager, REF: 180-03
Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from…
FDA · DeviceCovidienUS: AL, NJ, MA, MI, NY, NC, CO, TX, FL, KY, UT, CA, AZ, KS, IL, TN. OUS: El…Official notice →US
Icecure Medical Ltd recalls IceCure Cryoablation System ProSense -Intended for cryogenic destruction of tissue during surgical procedures by the application of…
Updated User Manual DSR3200000 Rev. E to include the safety guidelines in case of any mechanical damage (e.g. bent cryoprobe)
US
MEDLINE INDUSTRIES, LP - Northfield recalls Centurion Sterile CirClamp with 1.45cm Bell Reprocessed, Reorder CR3145K, 1 per package, 12 per case.
The kits incorrectly contain CirClamp subassembly 345CRSAK which results in the kits containing an incorrect bell and base size.
US
Philips Healthcare recalls Incisive CT, software version 5.0
Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted…
US
DKK Dai-Ichi Shomei Co., Ltd. recalls LFS Flat screen support arm systems without surgical lamp for models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4…
The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause…
US
ROi CPS LLC recalls GS0008 2Y - Burn Pack - STL, Item Number 880082025