Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and…
Read the official notice →- Date
- 2023-12-20
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
- Classification
- Class II
- Units
- US Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2024-05-01
US
Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature,…
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - US Nationwide including in the states of AL, AR, CA,…Official notice →- 2024-04-24
US
Siemens Medical Solutions USA, Inc recalls Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137…
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break .…
- 2024-01-17
US
Siemens Medical Solutions USA, Inc recalls Sensis Vibe Hemo
The possibility of the Sensis documentation functionality application to crash.
- 2023-11-22
US
Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)
Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…
- 2023-11-08
US
Siemens Medical Solutions USA, Inc recalls ***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and…
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
- 2023-10-11
US
Siemens Medical Solutions USA, Inc recalls (1) Over-ear headphones used with the MAGNETOM Amira MAGNETOM Free.Star MAGNETOM Sempra, MAGNETOM Free.Max, MAGNETOM Mica…
In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential…
Other drugs & medical recalls in the United States
- 2023-12-20
US
Encore Medical, LP recalls RSP Glenoid Baseplate, 30mm P2 Porous Coated REF 508-32-204. DJO Surgical shoulder device.
Coating hex did not engage with the Straight Mod Hexdriver or with the Hex Head Screw Driver
FDA · DeviceEncore Medical, LPU.S.: AZ, CA, FL, IN, KS, KY, ME, MN, MS, NJ, OH, Puerto Rico, RI, SC, TN, TX,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls STERILE PVP SOLUTION, REF DYNDA1907
The product description on the labeling includes the word "Sterile" or STRL", however the contents of the solution inside the foil sachet is non-sterile.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the states of CT, IA, LA, MA, PA, CA, NY, FL.Official notice →US
Limacorporate S.p.A recalls PHYSICA HPS Tibial Liner #6 H10 REF 6539.54.610
Due to incorrect product labeling
US
Medtronic Inc. recalls Medtronic SynchroMed, Model A10
The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus,…
US
Meridian Bioscience Inc recalls Cardinal Health Clean Stool Transportation System-For the collection, transportation, preservation, and examination of stool specimens…
Due to a manufacturing defect, certain Para-Pak vials may be at risk for leakage.
US
Meridian Bioscience Inc recalls Para-Pak Zn-PVA/10% Formalin-for the routine collection, transportation, preservation, and examination of stool specimens for intestinal…
Plastic vials contain a manufacturing defect may cause an increased risk of leakage and presents a risk of incidental exposure to biological hazards associated with direct contact with the patient specimen.