Siemens Medical Solutions USA, Inc recalls Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137…
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break .…
Read the official notice →- Date
- 2024-04-24
- Country
- United States
- Brand
- Siemens Medical Solutions
- Product
- Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee…
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Siemens Medical Solutions recalls
- 2024-07-31
US
Siemens Medical Solutions USA, Inc recalls Sensis Vibe systems with software version VD15B in combination with Sensis High End Server -Intended to be used as a…
If the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of CA, CT, FL, IA, KS, MA, MI, NJ,…Official notice →- 2024-07-03
US
Siemens Medical Solutions USA, Inc recalls AXIOM Vertix MD - The Vertix MD Trauma has been specially designed for examining emergency and accident patients as well as…
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
- 2024-05-01
US
Siemens Medical Solutions USA, Inc. recalls ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G
On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature,…
FDA · DeviceSiemens Medical Solutions USA, Inc.Worldwide distribution - US Nationwide including in the states of AL, AR, CA,…Official notice →- 2024-01-17
US
Siemens Medical Solutions USA, Inc recalls Sensis Vibe Hemo
The possibility of the Sensis documentation functionality application to crash.
- 2023-12-20
US
Siemens Medical Solutions USA, Inc recalls ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and…
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL…Official notice →- 2023-11-22
US
Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)
Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…
Other drugs & medical recalls in the United States
- 2024-05-01
US
US
Aruba Aloe Balm N.V. recalls Alcoholada Gel, Pain Relieving Gel, 0.5% Lidocaine Hydrochloride, packaged in (a) 8.5 fl oz (251 mL) plastic bottle UPC 0 82252 03120 9…
Chemical Contamination: Product manufactured with ethanol API that contains methanol
US
Pyramids Wholesale Inc. recalls Weiner Boner Honey, 12g packet, 100% Organic Formula.
Marketed without an approved NDA/ANDA: Products found to contain undeclared sildenafil and/or tadalafil.
- 2024-04-24
US
Smith & Nephew Medical, Ltd. recalls RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote…
Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump
FDA · DeviceSmith & Nephew Medical, Ltd.Worldwide distribution - US Nationwide and the countries of Canada, Chile.Official notice →US
Raye's Inc. recalls SW Bari Lift & Transfer, Model Number 38060000
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the…
FDA · DeviceRaye's Inc.Worldwide distribution - US Nationwide and the countries of Australia, UK.Official notice →US
Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample…