Raye's Inc. recalls SW Bari Lift & Transfer, Model Number 38060000
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the…
Read the official notice →- Date
- 2024-04-24
- Country
- United States
- Brand
- Raye's Inc.
- Product
- SW Bari Lift & Transfer, Model Number 38060000
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Australia, UK.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2024-05-01
US
Pyramids Wholesale Inc. recalls Weiner Boner Honey, 12g packet, 100% Organic Formula.
Marketed without an approved NDA/ANDA: Products found to contain undeclared sildenafil and/or tadalafil.
- 2024-04-24
US
Siemens Medical Solutions USA, Inc recalls Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137…
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break .…
US
Smith & Nephew Medical, Ltd. recalls RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote…
Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump
FDA · DeviceSmith & Nephew Medical, Ltd.Worldwide distribution - US Nationwide and the countries of Canada, Chile.Official notice →US
Randox Laboratories Ltd. recalls Microalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other methods and may misclassify patient results. Users may experience a shift in Quality Control and patient sample…
FDA · DeviceRandox Laboratories Ltd.US Nationwide distribution in the states of GA, MA, WV including PR.Official notice →US
ACTIM OY recalls Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane
Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.
US
Conformis, Inc. recalls Identity Imprint PS Tibial Tray Size 4: Lot 540287
Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.
FDA · DeviceConformis, Inc.US Nationwide distribution in the states of FL, KS and NH.Official notice →