Drugs & medicalClass II

Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient…

The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.

Read the official notice →
Date
2023-11-22
Country
United States
Brand
Ion Beam Applications
Product
IBA Proton Therapy System - PROTEUS 235. PTS-11 versions, PTS-12 versions with Orion PPS version OSS6.0. Configured with Orion Patient…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of Arkansas, Florida,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Ion Beam Applications recalls

  1. 2024-08-28
  2. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235- Designed to produce and deliver a proton beam for the treatment of patients with localized…

    Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues

    FDA · DeviceIon Beam Applications S.A.IL,LA, NJ, MI,OK, TN, TX, VA, WA Foreign: Czech Republic France Germany Italy…Official notice →
  3. 2024-06-12
  4. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized…

    Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances

    FDA · DeviceIon Beam Applications S.A.Worldwide - US Nationwide distribution in the state of Virginia and the…Official notice →
  5. 2024-05-01
  6. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls IBA Proton Therapy System - PROTEUS 235, PTS-10 versions, PTS-11 versions, and PTS-12 versions.

    There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.

    FDA · DeviceIon Beam Applications S.A.US Nationwide distribution in the states of AR, FL, LA, KS, MI, TX, VA.Official notice →
  7. 2022-06-08
  8. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other…

    First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose…

    FDA · DeviceIon Beam Applications S.A.Worldwide distribution - US Nationwide distribution in the states of IL, VA,…Official notice →
  9. 2022-03-16
  10. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and…

    When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually…

    FDA · DeviceIon Beam Applications S.A.IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England,…Official notice →
  11. 2022-01-19
  12. US
    Drugs & medicalClass II

    Ion Beam Applications S.A. recalls Proteus 235; Version: PTS-8 versions before PTS-8.7.2

    Proton Therapy System (PTS) software can be started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical…

    FDA · DeviceIon Beam Applications S.A.IN USA: VA, OK, NJ, TN OUS: Czech Republic.Official notice →
All 9 Ion Beam Applications recalls →

Other drugs & medical recalls in the United States

  1. 2023-11-22
  2. US
    Drugs & medicalClass II

    Microbiologics Inc recalls KWIKSTIK, P. venetum, Catalog numbers: a) 0794P (2-pack), b) 0794K (6-pack)

    A complaint investigation showed that the bulk pellets (lot 794-142) were contaminated with Aspergillus flavus.

    FDA · DeviceMicrobiologics IncWorldwide distribution - US Nationwide and the countries of Italy, France,…Official notice →
  3. US
    Drugs & medicalClass II

    Philips North America recalls Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

    When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.

    FDA · DevicePhilips North AmericaUS Nationwide distribution in the state of MN.Official notice →
  4. US
    Drugs & medicalClass II

    Physio-Control, Inc. recalls LIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051

    Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.

    FDA · DevicePhysio-Control, Inc.US Nationwide distribution in the states of AZ, CA, NC, NJ, NY.Official notice →
  5. US
    Drugs & medicalClass II

    Philips North America Llc recalls Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for radiation oncology and…

    For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect…

    FDA · DevicePhilips North America LlcNationwide Foreign: Austria Belgium Canada China Estonia France Germany…Official notice →
  6. US
    Drugs & medicalClass II

    Atrium Medical Corporation recalls Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac…

    Balloon-to-catheter weld failure that may cause loss of balloon pressure during stent deployment.

    FDA · DeviceAtrium Medical CorporationForeign: Australia, Belgium, Canada, France, Germany, Greece, Malta,…Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Mobilett Mira wireless (VA20) mobile x-ray system(s)

    Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual…

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide - WorldwideOfficial notice →
See all drugs & medical recalls →