Philips North America Llc recalls Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631
Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of…
Read the official notice →- Date
- 2023-01-18
- Country
- United States
- Brand
- Philips
- Product
- Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Australia,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2023-05-10
US
Philips North America Llc recalls Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and…
Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide including Puerto Rico and the countries…Official notice →- 2023-04-19
US
Philips North America Llc recalls Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through…
FDA · DevicePhilips North America LlcNationwide including Puerto Rico and Virgin Islands Foreign: ARGENTINA…Official notice →US
Philips North America Llc recalls CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…
- 2022-12-14
US
Philips North America Llc recalls Philips SmartPath to Ingenia Elition X MR System, Model Numbers 782118, 782132, 782144
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
- 2022-10-12
US
Philips North America Llc recalls MR Patient Care Portal 5000 (Desktop Unit), Model Number 453564792561 This is a component of product number 866162.
Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.
FDA · DevicePhilips North America LlcDomestic distribution to AL AZ CA CO FL GA ID IL IN LA MA MI MN NC NE NJ OH PA…Official notice →- 2022-09-07
US
Philips North America Llc recalls DigitalDiagnost C50 1.1
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in…
Other drugs & medical recalls in the United States
- 2023-01-18
US
Medtronic Neuromodulation recalls Vanta/Seqenita LT Clinician Programmer Application, Neurostimulation systems for pain therapy, Model Number CT900C
Software anomalies, CP App messages: "Too Many Device Found", "Unexpected Device Error Code 1502", and "System Update Needed, Service Code 303"
US
Bayer Medical Care, Inc. recalls Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affected material…
Black specks may be present on the packaged needles
FDA · DeviceBayer Medical Care, Inc.Worldwide distribution - US Nationwide distribution in the states of AZ, IN,…Official notice →US
Integra LifeSciences Corp. recalls Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457
Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever.
FDA · DeviceIntegra LifeSciences Corp.Worldwide distribution - US Nationwide distribution in the states of CA, CO,…Official notice →US
Ventana Medical Systems, Inc. recalls VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive)…
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent…
FDA · DeviceVentana Medical Systems, Inc.Worldwide distribution - U.S. Nationwide distribution including in the states…Official notice →US
Philips Respironics, Inc. recalls Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B,…
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially…
FDA · DevicePhilips Respironics, Inc.US Nationwide and Global Distribution to countries of: Australia, Brazil,…Official notice →US
Covidien LP recalls MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No.…
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
FDA · DeviceCovidien LPWorldwide distribution. US nationwide including Guam, USVI, and Puerto Rico;…Official notice →