Philips North America Llc recalls MR Patient Care Portal 5000 (Desktop Unit), Model Number 453564792561 This is a component of product number 866162.
Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.
Read the official notice →- Date
- 2022-10-12
- Country
- United States
- Brand
- Philips
- Product
- MR Patient Care Portal 5000 (Desktop Unit), Model Number 453564792561 This is a component of product number 866162.
- Classification
- Class II
- Units
- Domestic distribution to AL AZ CA CO FL GA ID IL IN LA MA MI MN NC NE NJ OH PA…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2023-04-19
US
Philips North America Llc recalls CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…
- 2023-01-18
US
Philips North America Llc recalls Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631
Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2022-12-14
US
Philips North America Llc recalls Philips SmartPath to Ingenia Elition X MR System, Model Numbers 782118, 782132, 782144
Gradient Coil may act as a heat source with a potential to produce smoke and/or fire.
- 2022-09-07
US
Philips North America Llc recalls DigitalDiagnost C50 1.1
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in…
- 2022-05-18
US
Philips North America Llc recalls Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of…
When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure…
FDA · DevicePhilips North America LlcDevices were distributed to the following US states: FL, ID, NJ, OK, SC and…Official notice →- 2022-04-20
US
Philips North America, LLC recalls The IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for the following product model #s:…
Update to instructions for use regarding the cleaning and disinfection process for the reusable probes.
FDA · DevicePhilips North America, LLCWorldwide distribution: US (nationwide) Including Puerto Rico and OUS…Official notice →
Other drugs & medical recalls in the United States
- 2022-10-12
US
Synthes (USA) Products LLC recalls Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate…
US
Zimmer, Inc. recalls G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically…
US
Wright Medical Technology, Inc. recalls stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System…
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm…
US
US
Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0
The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does…
US
AuroMedics Pharma LLC recalls Tranexamic Acid Injection, USP, 1000mg per 10 mL (100mg /mL), 10mL single-dose vial, Rx Only, Distributed by: AuroMedics Pharma LLC E.…
Presence of Particulate Matter: Piece of metal found in a vial