Ventana Medical Systems, Inc. recalls VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive)…
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent…
Read the official notice →- Date
- 2023-01-18
- Country
- United States
- Brand
- Ventana Medical Systems
- Product
- VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive)…
- Classification
- Class II
- Units
- Worldwide distribution - U.S. Nationwide distribution including in the states…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Ventana Medical Systems recalls
- 2026-08-19
US
Ventana Medical Systems, Inc. recalls navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number…
Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window
FDA · DeviceVentana Medical Systems, Inc.Worldwide - U.S. Nationwide distribution including in the states of AL, AR,…Official notice →- 2024-08-28
US
Ventana Medical Systems, Inc. recalls Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
FDA · DeviceVentana Medical Systems, Inc.Worldwide - US Nationwide distribution including in the states of AK, AL, AR,…Official notice →- 2022-07-20
US
Ventana Medical Systems Inc recalls Ventana HE 600 System, automated slide preparer, for use in laboratories.
There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
FDA · DeviceVentana Medical Systems IncUS: AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD MI MN MO MS MT NC NE NJ…Official notice →- 2022-04-13
US
Ventana Medical Systems Inc recalls BenchMark ULTRA and DISCOVERY ULTRA Instruments
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
FDA · DeviceVentana Medical Systems IncUS: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND,…Official notice →
Other drugs & medical recalls in the United States
- 2023-01-18
US
Bayer Medical Care, Inc. recalls Source Administration Set (SAS) used with the Medrad Intego PET Infusion System, Sterile Catalog number: INT CSS Affected material…
Black specks may be present on the packaged needles
FDA · DeviceBayer Medical Care, Inc.Worldwide distribution - US Nationwide distribution in the states of AZ, IN,…Official notice →US
Integra LifeSciences Corp. recalls Codman Surgical Patties, Part number 801400; SURG PATXRAY 1/2X1/2-200; Lot 6396457
Integra LifeSciences identified that one lot was released with out-of-specification endotoxin results. The worst-case risk for the use of affected Surgical Patties is potentially a minor transient fever.
FDA · DeviceIntegra LifeSciences Corp.Worldwide distribution - US Nationwide distribution in the states of CA, CO,…Official notice →US
Philips North America Llc recalls Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631
Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Philips Respironics, Inc. recalls Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B,…
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially…
FDA · DevicePhilips Respironics, Inc.US Nationwide and Global Distribution to countries of: Australia, Brazil,…Official notice →US
Covidien LP recalls MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No.…
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
FDA · DeviceCovidien LPWorldwide distribution. US nationwide including Guam, USVI, and Puerto Rico;…Official notice →US
Emergent Protective Products USA Inc recalls RSDL (Reactive Skin Decontamination Lotion) Kit, NSN 6505-21-912-5229. Used to remove and/or neutralize chemical warfare…
Complaints of leaking packets were received.
FDA · DeviceEmergent Protective Products USA IncDistribution to US state of VA (distributor), Canada, and SwitzerlandOfficial notice →