Covidien LP recalls MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No.…
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Read the official notice →- Date
- 2023-01-18
- Country
- United States
- Brand
- Covidien
- Product
- MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No.…
- Classification
- Class I
- Units
- Worldwide distribution. US nationwide including Guam, USVI, and Puerto Rico;…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Covidien recalls
- 2023-03-08
US
Covidien, LP recalls Parietex Composite Mesh Polyester with Absorbable Collagen Film, Horseshoe- Shaped, 9 x 8 cm (3.6" x 3.1"). Intended for the reinforcement…
Collagen film was placed on the opposite side of the mesh may cause delay to treatment/therapy, adhesions, erosion/migration, pain, fistula, hernia (recurrence), and failure of implant
FDA · DeviceCovidien, LPUS Distribution to states of: CA, GA, NY and OUS Foreign distribution to…Official notice →US
Covidien, LP recalls Surgilon Braided Nylon suture Product Description: 88861883-51 SURGILON* 2-0 BLK 75CM C16 88861915-71 SURGILON* 1 BLK 75CM GS11…
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
FDA · DeviceCovidien, LPUS Nationwide Distribution and OUS Foreign distribution to countries of:…Official notice →- 2023-01-18
US
Covidien, LLC recalls COVIDIEN Emprint Ablation Catheter with Thermosphere Technology, Bronchoscopic Ablation, REF CA108L1
The product is labeled with the incorrect expiration date.
- 2022-11-23
US
Covidien, LP recalls Endo Stitch V-Loc 180 Absorbable Reload, Models VLOCA004L, VLOCA006L, VLOCA008L, VLOCA204L, VLOCA206L, VLOCA208L, VLOCA304L, VLOCA306L,…
Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign…
FDA · DeviceCovidien, LPDomestic distribution nationwide. Foreign distribution to Australia Austria…Official notice →- 2022-09-28
US
Covidien Llc recalls Puritan Bennett 560 Ventilator, PB560, CFN 4096600
Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.
FDA · DeviceCovidien LlcDomestic distribution to one customer in Washington, D.C. Foreign distribution…Official notice →- 2022-08-03
US
Covidien Llc recalls TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number:…
Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.
FDA · DeviceCovidien LlcAZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom…Official notice →
Other drugs & medical recalls in the United States
- 2023-01-18
US
Philips North America Llc recalls Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631
Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Ventana Medical Systems, Inc. recalls VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive)…
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent…
FDA · DeviceVentana Medical Systems, Inc.Worldwide distribution - U.S. Nationwide distribution including in the states…Official notice →US
Philips Respironics, Inc. recalls Philips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B,…
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially…
FDA · DevicePhilips Respironics, Inc.US Nationwide and Global Distribution to countries of: Australia, Brazil,…Official notice →US
Emergent Protective Products USA Inc recalls RSDL (Reactive Skin Decontamination Lotion) Kit, NSN 6505-21-912-5229. Used to remove and/or neutralize chemical warfare…
Complaints of leaking packets were received.
FDA · DeviceEmergent Protective Products USA IncDistribution to US state of VA (distributor), Canada, and SwitzerlandOfficial notice →US
US
Direct Rx recalls Moxifloxacin Ophthalmic Solution, 0.5%, 3 mL, Rx Only, Packaged and Distributed By: Direct Rx, LLC 94 Worldwide Drive, Dawsonville, GA…