Drugs & medicalClass II

Covidien, LP recalls Endo Stitch V-Loc 180 Absorbable Reload, Models VLOCA004L, VLOCA006L, VLOCA008L, VLOCA204L, VLOCA206L, VLOCA208L, VLOCA304L, VLOCA306L,…

Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign…

Read the official notice →
Date
2022-11-23
Country
United States
Brand
Covidien
Product
Endo Stitch V-Loc 180 Absorbable Reload, Models VLOCA004L, VLOCA006L, VLOCA008L, VLOCA204L, VLOCA206L, VLOCA208L, VLOCA304L, VLOCA306L,…
Classification
Class II
Units
Domestic distribution nationwide. Foreign distribution to Australia Austria…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Covidien recalls

  1. 2023-03-08
  2. US
    Drugs & medicalClass II

    Covidien, LP recalls Surgilon Braided Nylon suture Product Description: 88861883-51 SURGILON* 2-0 BLK 75CM C16 88861915-71 SURGILON* 1 BLK 75CM GS11…

    Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

    FDA · DeviceCovidien, LPUS Nationwide Distribution and OUS Foreign distribution to countries of:…Official notice →
  3. 2023-01-18
  4. US
    Drugs & medicalClass II

    Covidien, LLC recalls COVIDIEN Emprint Ablation Catheter with Thermosphere Technology, Bronchoscopic Ablation, REF CA108L1

    The product is labeled with the incorrect expiration date.

    FDA · DeviceCovidien, LLCForeign Distribution: Hong KongOfficial notice →
  5. US
    Drugs & medicalClass I

    Covidien LP recalls MAHURKAR 13.5Fr High Flow Dual Lumen Acute Dialysis Catheter, Sterile, Single Use, Rx Only: 13.5cm, Straight Extensions, Kit - Model No.…

    There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

    FDA · DeviceCovidien LPWorldwide distribution. US nationwide including Guam, USVI, and Puerto Rico;…Official notice →
  6. 2022-09-28
  7. US
    Drugs & medicalClass II

    Covidien Llc recalls Puritan Bennett 560 Ventilator, PB560, CFN 4096600

    Medtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by manufacturing error of specific turbine components.

    FDA · DeviceCovidien LlcDomestic distribution to one customer in Washington, D.C. Foreign distribution…Official notice →
  8. 2022-08-03
  9. US
    Drugs & medicalClass II

    Covidien Llc recalls TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number:…

    Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.

    FDA · DeviceCovidien LlcAZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom…Official notice →
  10. 2022-07-13
  11. US
    Drugs & medicalClass II

    Covidien, LP recalls EEA Autosuture Circular Stapler with DST Series Technology, 25mm. Model Numbers EEA25, EEAXL25, EEA2535, and EEAXL2535.

    The affected staplers have the potential for the staple guide to not be securely attached to the instrument. The issue could cause the component to disengage and if disengaged, could allow the device to transect tissue…

    FDA · DeviceCovidien, LPWorldwide distribution - US Nationwide.Official notice →
All 53 Covidien recalls →

Other drugs & medical recalls in the United States

  1. 2022-11-23
  2. US
    Drugs & medicalClass II

    Spectranetics Corporation recalls Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the cardiovascular system, and for the…

    Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation…

    FDA · DeviceSpectranetics CorporationUS: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD,…Official notice →
  3. US
    Drugs & medicalClass II

    Carefusion 2200 Inc recalls CareFusion V.Mueller GENESIS FULL-LENGTH CONTAINER 23 X12" (58.4 X 30.5CM) 6" (15.2CM) DEEP STERRAD COMPATIBLE, REF CD3-6ST

    Genesis STERRAD Reusable Rigid Sterilization Containers have not consistently met the requirements for aerosol challenge testing.

    FDA · DeviceCarefusion 2200 IncUS and Australia, Bahrain, Canada, Denmark, Hong Kong, Italy, Japan, Malaysia,…Official notice →
  4. US
    Drugs & medicalClass III

    Canadian Hospital Specialties recalls MED-RX UROLOGICAL CYSTO/BLADDER IRRIGATION SET, REF 10-4001

    Some of the labels of the kits for product code 10-4001 in the specified lots have an illegible barcode which prevents successful scanning of the package.

    FDA · DeviceCanadian Hospital SpecialtiesUSOfficial notice →
  5. US
    Drugs & medicalClass I

    Insulet Corporation recalls 18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM…

    The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a…

    FDA · DeviceInsulet CorporationWorldwide Distribution: US nationwide, Canada, Europe, Australia, and the…Official notice →
  6. US
    Drugs & medicalClass II

    Adam's Polishes LLC recalls Adam's Polishes ALCOHOL BASED HAND SANITIZER, isopropyl alcohol 75% v/v, packaged in a) 4 fl. oz (118) spray bottle, NDC 74943-125-04; b)…

    CGMP Deviations: Other lots of hand sanitizer are being recalled because they were manufactured under the same conditions as the product lot found to contain methanol.

    FDA · DrugAdam's Polishes LLCNationwide in the USAOfficial notice →
  7. US
    Drugs & medicalClass I

    Adam's Polishes LLC recalls Adam's Polishes ALCOHOL BASED HAND SANITIZER, isopropyl alcohol 75% v/v, packaged in a) 4 fl. oz (118) spray bottle, NDC 74943-125-04; b)…

    Chemical Contamination: FDA analysis found 1 lot of Adam's Polishes ALCOHOL BASED HAND SANITIZER isopropyl alcohol 75% v/v to contain methanol.

    FDA · DrugAdam's Polishes LLCNationwide in the USAOfficial notice →
See all drugs & medical recalls →