Philips North America Llc recalls Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and…

Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop…

Read the official notice →
Date
2023-05-10
Country
United States
Brand
Philips
Product
Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and…
Classification
Class II
Units
Worldwide distribution - US Nationwide including Puerto Rico and the countries…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips recalls

  1. 2023-05-24
  2. US
    Drugs & medicalClass II

    Philips North America recalls ProxiDiagnost N90 R.1.0

    Philips has identified a compliance issue where ProxiDiagnost N90 R.1.0, CombiDiagnost R90 R.1.0 and CombiDiagnost R90 R1.1 systems are missing a certification label required by the Food and Drug Administration (FDA).

    FDA · DevicePhilips North AmericaUS NationwideOfficial notice →
  3. US
    Drugs & medicalClass II

    Philips North America recalls Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813

    Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury

    FDA · DevicePhilips North AmericaWorldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN,…Official notice →
  4. US
    Drugs & medicalClass II

    Philips North America Llc recalls Spectral CT 7500 Computed Tomography X-Ray System, Model Number 728333

    Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operators foot. If foot or leg entrapment occurs while pressing Unload Pedal, the risk to operators may include:…

    FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution including in the states of States: AZ,…Official notice →
  5. 2023-04-19
  6. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips DigitalDiagnost C50 -intended for use in generating radiographic images of human anatomy Model Number : 712201

    If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through…

    FDA · DevicePhilips North America LlcNationwide including Puerto Rico and Virgin Islands Foreign: ARGENTINA…Official notice →
  7. US
    Drugs & medicalClass II

    Philips North America Llc recalls CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…

    FDA · DevicePhilips North America LlcUS Nationwide. Global Distribution.Official notice →
  8. 2023-01-18
  9. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Fetal Spiral Electrode-intended for patients requiring fetal heart rate monitoring during labor. PN: 989803137631

    Fetal Spiral Electrode (FSE) spiral tip of the FSE breaking off during use and requiring surgical intervention to remove the broken tip from the neonate patient. Philips has decided to discontinue the distribution of…

    FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →
All 100 Philips recalls →

Other drugs & medical recalls in the United States

  1. 2023-05-10
  2. US
    Drugs & medicalClass II

    Beaver Visitec International, Inc. recalls CryoTreQ- Aa disposable handheld instrument intended for ophthalmic surgery. It creates a tip at cryogenic temperatures by…

    Device suddenly and forcefully disassembled after cryotherapy, potential for long-term or permanent impairment of a patient or user if a CryoTreq device suddenly or forcefully disassembles during a cryosurgery procedure.

    FDA · DeviceBeaver Visitec International, Inc.US Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Hill-Rom pro+ mattress, product codes (REF): a) 20877002S, b) 20877003S, c) 20877004S, d) 20877007S, e) 20877008S, f) 21649301S, g)…

    There is a potential for mattress cover damage (delamination) which may result in fluid absorption in the mattress (fluid ingress).

    FDA · DeviceBaxter Healthcare CorporationWorldwide distribution - US Nationwide and the countries of Canada, Qatar.Official notice →
  4. US
    Drugs & medicalClass II

    Labories Medical Technologies recalls injeTAK Adjustable Tip Needles Needle Sheath diameter 6Fr with 35-50cm in length Stainless-stell needle cannula is 25 gauge

    The packaging may be damaged, compromising the sterile barrier.

    FDA · DeviceLabories Medical TechnologiesWorldwide distribution.Official notice →
  5. US
    Drugs & medicalClass II

    Steris Corporation recalls Steris Lighting and Visualization Systems, Sterile Light Handle Cover (LB53)- Surgical light handle covers, including disposable surgical…

    Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient

    FDA · DeviceSteris CorporationWorldwide - US Nationwide distribution including Puerto Rico and the countries…Official notice →
  6. US
    Drugs & medicalClass II

    Stryker Corporation recalls SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200

    Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.

    FDA · DeviceStryker CorporationWorldwide - US Nationwide distribution including in the states of PA, MS, IL,…Official notice →
  7. US
    Drugs & medicalClass II

    Ottlite TEchnologies, Inc. recalls OttLite Rechargeable UVC Disinfecting Wand

    Product allows for the emission of ultraviolet (UV) electronic product radiation in excess of exposure limits recommended by the International Commission on Non-ionizing Radiation Protection (ICNIRP). The emissions are…

    FDA · DeviceOttlite TEchnologies, Inc.US Natioonwide DistributionOfficial notice →
See all drugs & medical recalls →